Electrocardiograms with pacemakers: accuracy of computer reading
- PMID: 16919672
- DOI: 10.1016/j.jelectrocard.2006.07.001
Electrocardiograms with pacemakers: accuracy of computer reading
Abstract
Objective: We analyzed the accuracy with which a computer algorithm reads electrocardiograms (ECGs) with electronic pacemakers (PMs).
Methods: Electrocardiograms were screened for the presence of electronic pacing spikes. Computer-derived interpretations were compared with cardiologists' readings.
Results: Computer-drawn interpretations required revision by cardiologists in 61.3% of cases. In 18.4% of cases, the ECG reading algorithm failed to recognize the presence of a PM. The misinterpretation of paced beats as intrinsic beats led to multiple secondary errors, including myocardial infarctions in varying localization. The most common error in computer reading was the failure to identify an underlying rhythm. This error caused frequent misidentification of the PM type, especially when the presence of normal sinus rhythm was not recognized in a tracing with a DDD PM tracking the atrial activity.
Conclusion: The increasing number of pacing devices, and the resulting number of ECGs with pacing spikes, mandates the refining of ECG reading algorithms. Improvement is especially needed in the recognition of the underlying rhythm, pacing spikes, and mode of pacing.
Similar articles
-
Accuracy of electrocardiogram interpretation by cardiologists in the setting of incorrect computer analysis.J Electrocardiol. 2006 Jul;39(3):343-5. doi: 10.1016/j.jelectrocard.2006.02.002. J Electrocardiol. 2006. PMID: 16777525
-
Errors in the computerized electrocardiogram interpretation of cardiac rhythm.J Electrocardiol. 2007 Sep-Oct;40(5):385-90. doi: 10.1016/j.jelectrocard.2007.03.008. Epub 2007 May 24. J Electrocardiol. 2007. PMID: 17531257
-
Common errors in computer electrocardiogram interpretation.Int J Cardiol. 2006 Jan 13;106(2):232-7. doi: 10.1016/j.ijcard.2005.02.007. Int J Cardiol. 2006. PMID: 16321696
-
Modern pacemakers: hope or hype?Pacing Clin Electrophysiol. 2009 Sep;32(9):1207-21. doi: 10.1111/j.1540-8159.2009.02467.x. Pacing Clin Electrophysiol. 2009. PMID: 19719501 Review.
-
The Computerized ECG: Friend and Foe.Am J Med. 2019 Feb;132(2):153-160. doi: 10.1016/j.amjmed.2018.08.025. Epub 2018 Sep 8. Am J Med. 2019. PMID: 30205084 Review.
Cited by
-
A simple, wide bandwidth, biopotential amplifier to record pacemaker pulse waveform.Med Devices (Auckl). 2016 Sep 16;9:325-329. doi: 10.2147/MDER.S97902. eCollection 2016. Med Devices (Auckl). 2016. PMID: 27695369 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials