Evaluation of the anterior chamber angle in Asian Indian eyes by ultrasound biomicroscopy and gonioscopy

Affiliations
Free article
Comparative Study

Evaluation of the anterior chamber angle in Asian Indian eyes by ultrasound biomicroscopy and gonioscopy

Sushmita Kaushik et al. Indian J Ophthalmol. 2006 Sep.
Free article

Abstract

Purpose: To compare the ultrasound biomicroscopic measurement of the anterior chamber angle in Asian Indian eyes, with the angle width estimated by gonioscopy.

Participants: Patients with open and closed angles attending a glaucoma clinic were recruited for the study.

Observation procedures: Temporal quadrants of the angles of patients were categorized by gonioscopy as Grade 0 to Grade 4, using Shaffer's classification. These angles were quantified by ultrasound biomicroscopy (UBM) using the following biometric characteristics: Angle opening distance at 250 micro (AOD 250) and 500 micro (AOD 500) from the scleral spur and trabecular meshwork-ciliary process distance (TCPD). The angles were further segregated as "narrow angles" (Schaffer's Grade 2 or less) and "open angles" (Schaffer's Grade 3 and 4).

Main outcome measures: The UBM measurements were computed in each case and analyzed in relation to the gonioscopic angle evaluation.

Results: One hundred and sixty three eyes of 163 patients were analyzed. One hundred and six eyes had "narrow angles" and 57 eyes had "open angles" on gonioscopy. There was a significant difference among the mean UBM measurements of each angle grade estimated by gonioscopy (P < 0.001). The Pearson correlation coefficient between all UBM parameters and gonioscopy grades was significant at the 0.01 level. The mean AOD 250, AOD 500 and TCPD in narrow angles were 58+/-49 micro, 102+/-84 micro and 653+/-124 respectively, while it was 176+/-47 micro, 291+/-62 micro and 883+/-94 micro in eyes with open angles (P < 0.001) respectively.

Conclusions: The angle width estimated by gonioscopy correlated significantly with the angle dimensions measured by UBM. Gonioscopy, though a subjective test, is a reliable method for estimation of the angle width.