Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2006 Sep;82(3):917-24; discussion 924-5.
doi: 10.1016/j.athoracsur.2006.03.065.

Improvement in survival after mechanical circulatory support with pneumatic pulsatile ventricular assist devices in pediatric patients

Affiliations
Comparative Study

Improvement in survival after mechanical circulatory support with pneumatic pulsatile ventricular assist devices in pediatric patients

Roland Hetzer et al. Ann Thorac Surg. 2006 Sep.

Abstract

Background: Pediatric size pneumatically driven extracorporeal ventricular assist devices (VAD) for infants and small children were introduced into clinical routine in 1992. In the initial period, the results in infants were poor. Since then, several improvements have been introduced with regard to the cannulas, connectors, heparin coating of the blood pump inner surface, anticoagulant treatment and coagulation monitoring, and earlier decision-making in favor of pump implantation before irreversible shock has set in.

Methods: Since 1990 and as of January 1, 2005, 62 Berlin Heart Excor systems have been implanted in patients below 18 years of age at our institution. The patients were divided into two groups according to the period of treatment: period 1, devices implanted between 1990 and 1998 (n = 34), and period 2, devices implanted between 1999 and 2004 (n = 28). We compared our experience during the earlier and later periods.

Results: There were no significant differences in the preoperative patient data between the two periods except for time of support (17.9 +/- 27.7 days versus 53.2 +/- 83.9 days, p = 0.001). In period 1, more patients needed a biventricular VAD whereas in period 2, more patients were effectively treated with a left VAD (p = 0.05). In the later period, the chest could be primarily closed in a significantly higher percentage of infants (0% versus 89%, p = 0.012), and more infants could be extubated on the VAD (0% versus 55%, p = 0.16). Discharge from the hospital after either weaning from the system or heart transplantation was achieved for 35% in period 1 and for 68% in period 2 (p = 0.029). Whereas in period 1 there were no survivors in the group of children younger than 1 year old, during period 2, survival in this age group was similar to that of the two groups of older children (p = 0.024). There was a significant improvement in the discharge rate in period 2 in patients with cardiomyopathy (43% versus 76%, p = 0.045) and postcardiotomy heart failure (0% versus 57%, p = 0.01).

Conclusions: Earlier implantation of VADs, heparin coating of the blood pumps, and substantial modifications in cannula design, anticoagulation, and the coagulation monitoring regimen have led to a significant increase in the survival and discharge rate, especially among children under 1 year of age. The pediatric size Berlin Heart Excor VAD is a valuable option as a bridge to heart transplantation or recovery for children suffering from cardiogenic shock.