. 2007 Jun;29(5):580-5.
doi: 10.1016/j.medengphy.2006.06.012. Epub 2006 Aug 22.

An instrumented implant for vertebral body replacement that measures loads in the anterior spinal column

Affiliations

An instrumented implant for vertebral body replacement that measures loads in the anterior spinal column

Antonius Rohlmann et al. Med Eng Phys. 2007 Jun.

Abstract

Realistic loads on a spinal implant are required among others for optimization of implant design and preclinical testing. In addition, such data may help to choose the optimal physiotherapy program for patients with such an implant and to evaluate the efficacy of aids like braces or crutches. Presently, no implant is available that can measure loads in the anterior spinal column during activities of daily life. Therefore, an implant instrumented for in vivo load measurement was developed for vertebral body replacement. The aim of this paper is to describe in detail a telemeterized implant that measures forces and moments acting on it. Six load sensors, a nine-channel telemetry unit and a coil for inductive power supply of the electronic circuits were integrated into a modified vertebral body replacement (Synex). The instrumented part of the implant is hermetically sealed. Patients are videotaped during measurements, and implant loads are displayed on and off line. The average accuracy of load measurement is better than 2% for force and 5% for moment components with reference to the maximum value of 3000 N and 20 Nm, respectively. The measuring implant described here will provide additional information on spinal loads.