Three-dimensional myocardial activation imaging in a rabbit model
- PMID: 16941837
- DOI: 10.1109/TBME.2006.873701
Three-dimensional myocardial activation imaging in a rabbit model
Abstract
This study evaluated a recently developed three-dimensional (3-D) electrocardiographic imaging (3DECI) approach in a closed-chest rabbit model. First, the performance and sensitivity of parameters of 3DECI were evaluated using a geometrically realistic rabbit heart-torso model. Second, a 3-D intracardiac mapping procedure was evaluated based on the heart-torso rabbit model. Third, comparisons were made among the forward-simulated ventricular activation sequence, the activation sequence derived by the intracardiac mapping, and the 3DECI inverse solution. Finally, the present procedures were tested in vivo in a rabbit, in which the relative error between the measured and imaged activation sequence was 0.20 and the localization error was 5.1 mm. The present simulation and experimental results suggest the merits of the 3DECI imaging approach, and the validity of intracardiac mapping as a tool to evaluate the 3DECI.
Similar articles
-
Wavefront-based models for inverse electrocardiography.IEEE Trans Biomed Eng. 2006 Sep;53(9):1821-31. doi: 10.1109/TBME.2006.878117. IEEE Trans Biomed Eng. 2006. PMID: 16941838
-
Use of group theory in the selection and description of regularization methods for functional source imaging.IEEE Trans Biomed Eng. 2006 Sep;53(9):1832-40. doi: 10.1109/TBME.2006.873693. IEEE Trans Biomed Eng. 2006. PMID: 16941839
-
Three-dimensional cardiac electrical imaging from intracavity recordings.IEEE Trans Biomed Eng. 2007 Aug;54(8):1454-60. doi: 10.1109/TBME.2007.891932. IEEE Trans Biomed Eng. 2007. PMID: 17694866
-
3D Heart: a new visual training method for electrocardiographic analysis.J Electrocardiol. 2007 Sep-Oct;40(5):457.e1-7. doi: 10.1016/j.jelectrocard.2007.04.001. Epub 2007 Jul 2. J Electrocardiol. 2007. PMID: 17604044 Review.
-
Selecting CRT candidates: the value of intracardiac mapping.Europace. 2008 Nov;10 Suppl 3:iii106-9. doi: 10.1093/europace/eun224. Europace. 2008. PMID: 18955390 Review.
Cited by
-
Noninvasive imaging of cardiac electrophysiology.Indian Pacing Electrophysiol J. 2007 Aug 1;7(3):160-5. Indian Pacing Electrophysiol J. 2007. PMID: 17684574 Free PMC article.
-
Noninvasive three-dimensional cardiac activation imaging from body surface potential maps: a computational and experimental study on a rabbit model.IEEE Trans Med Imaging. 2008 Nov;27(11):1622-30. doi: 10.1109/TMI.2008.929094. IEEE Trans Med Imaging. 2008. PMID: 18955177 Free PMC article.
-
Estimation of global ventricular activation sequences by noninvasive three-dimensional electrical imaging: validation studies in a Swine model during pacing.J Cardiovasc Electrophysiol. 2008 May;19(5):535-40. doi: 10.1111/j.1540-8167.2007.01066.x. Epub 2007 Dec 20. J Cardiovasc Electrophysiol. 2008. PMID: 18179521 Free PMC article.
-
Temporal Sparse Promoting Three Dimensional Imaging of Cardiac Activation.IEEE Trans Med Imaging. 2015 Nov;34(11):2309-19. doi: 10.1109/TMI.2015.2429134. Epub 2015 May 4. IEEE Trans Med Imaging. 2015. PMID: 25955987 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical