Preliminary kinematic evaluation of a new stance-control knee-ankle-foot orthosis
- PMID: 16949186
- DOI: 10.1016/j.clinbiomech.2006.06.008
Preliminary kinematic evaluation of a new stance-control knee-ankle-foot orthosis
Abstract
Background: Stance-control knee-ankle-foot orthoses permit free knee motion in swing while providing knee flexion resistance in stance for individuals with quadriceps muscle weakness. However, some stance-control knee-ankle-foot orthoses require full knee extension to engage the knee-joint lock, thereby not providing knee support when climbing stairs or stepping over curbs. Stance-control knee-ankle-foot orthoses that do support a flexed knee are either heavy, bulky, expensive, offer a limited number of locking positions, or cause noise. This paper presents a preliminary kinematic evaluation of a new stance-control knee-ankle-foot orthosis that was designed to address these limitations.
Methods: Kinematic gait analysis was performed on three male knee-ankle-foot-orthosis users with knee extensor weakness in at least one limb (mean age: 56.3 years (SD 4.0)). Three walking trials were performed with the subjects' current knee-ankle-foot-orthosis and then the new stance-control knee-ankle-foot orthosis (non-randomized before-after trial). Subjects completed a questionnaire about the new stance-control knee-ankle-foot orthosis and current knee-ankle-foot-orthosis.
Findings: A mean increase in knee flexion of 21.1 degrees (SD 8.2) during swing and a greater total knee range of motion was found when walking with the new stance-control knee-ankle-foot orthosis. Two knee-ankle-foot-orthosis users experienced a reduction in pelvic obliquity and hip abduction angle abnormalities when walking with the stance-control knee-ankle-foot orthosis. Two out of three subjects preferred walking with the new stance-control knee-ankle-foot orthosis over their prescribed knee-ankle-foot-orthosis.
Interpretation: The new stance-control knee-ankle-foot orthosis permitted improved gait kinematics for knee-ankle-foot-orthosis users while providing knee support in stance and free knee motion in swing at appropriate instants in the gait cycle. Overall, the new stance-control knee-ankle-foot orthosis provided more natural gait kinematics for orthosis users compared to conventional knee-ankle-foot-orthoses.
Similar articles
-
The effect of stance control orthoses on gait characteristics and energy expenditure in knee-ankle-foot orthosis users.Prosthet Orthot Int. 2010 Jun;34(2):206-15. doi: 10.3109/03093641003773189. Prosthet Orthot Int. 2010. PMID: 20470059
-
Engineering design review of stance-control knee-ankle-foot orthoses.J Rehabil Res Dev. 2009;46(2):257-67. J Rehabil Res Dev. 2009. PMID: 19533539
-
Design and evaluation of a stance-control knee-ankle-foot orthosis knee joint.IEEE Trans Neural Syst Rehabil Eng. 2006 Sep;14(3):361-9. doi: 10.1109/TNSRE.2006.881578. IEEE Trans Neural Syst Rehabil Eng. 2006. PMID: 17009496
-
The gait and energy efficiency of stance control knee-ankle-foot orthoses: A literature review.Prosthet Orthot Int. 2016 Apr;40(2):202-14. doi: 10.1177/0309364615588346. Epub 2015 Jun 8. Prosthet Orthot Int. 2016. PMID: 26055252 Review.
-
The impact of ankle-foot orthosis stiffness on gait: A systematic literature review.Gait Posture. 2019 Mar;69:101-111. doi: 10.1016/j.gaitpost.2019.01.020. Epub 2019 Jan 15. Gait Posture. 2019. PMID: 30708092
Cited by
-
Safety and walking ability of KAFO users with the C-Brace® Orthotronic Mobility System, a new microprocessor stance and swing control orthosis.Prosthet Orthot Int. 2017 Feb;41(1):65-77. doi: 10.1177/0309364616637954. Epub 2016 Jul 10. Prosthet Orthot Int. 2017. PMID: 27151648 Free PMC article.
-
Use and Usability Of Custom-Made Knee-Ankle-Foot Orthoses In Polio Survivors with Knee Instability: A Cross-Sectional Survey.J Rehabil Med. 2022 Feb 14;54:jrm00261. doi: 10.2340/jrm.v53.1122. J Rehabil Med. 2022. PMID: 34931255 Free PMC article.
-
Biomechanical effect of electromechanical knee-ankle-foot-orthosis on knee joint control in patients with poliomyelitis.Med Biol Eng Comput. 2008 Jun;46(6):541-9. doi: 10.1007/s11517-008-0310-6. Epub 2008 Feb 8. Med Biol Eng Comput. 2008. PMID: 18259793
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical