Impact of electrode insertion depth on intracochlear trauma
- PMID: 16949967
- DOI: 10.1016/j.otohns.2006.05.002
Impact of electrode insertion depth on intracochlear trauma
Abstract
Objective: To assess the effect of cochlear implant (CI) insertion depth and surgical technique on intracochlear trauma.
Study design and setting: Twenty-one fresh human temporal bones were implanted with CI electrodes and underwent histologic processing and evaluation. Specimens were grouped into 3 categories: 1) soft implantation technique and standard electrode; 2) soft implantation technique and flexible prototype array; 3) forceful implantations and standard electrode. Based on the grading system (1 to 4), 2 numeric values were calculated indicating the overall severity of cochlear damage (trauma indices).
Results: Mean trauma index values were 13.8, 36.3, and 59.2 for group 1, 2, and 3, respectively. Differences in cochlear trauma (trauma index) were nonsignificant between specimens in groups 1 and 2 but were significant between groups 1 and 3.
Conclusion: This study gives evidence that intracochlear trauma increases with deep insertions. Thus, in cases where cochlear integrity might be important, limited insertions should be achieved.
Similar articles
-
[Evaluation of the insertion-trauma of the Nucleus Contour Advance electrode-array in a human temporal bone model].Laryngorhinootologie. 2004 Dec;83(12):840-4. doi: 10.1055/s-2004-826067. Laryngorhinootologie. 2004. PMID: 15611904 German.
-
Combining perimodiolar electrode placement and atraumatic insertion properties in cochlear implantation -- fact or fantasy?Acta Otolaryngol. 2006 May;126(5):475-82. doi: 10.1080/00016480500437393. Acta Otolaryngol. 2006. PMID: 16698696
-
Insertional trauma of multichannel cochlear implants.Laryngoscope. 1993 Sep;103(9):995-1001. doi: 10.1288/00005537-199309000-00010. Laryngoscope. 1993. PMID: 8361322
-
Cochleostomy site: implications for electrode placement and hearing preservation.Acta Otolaryngol. 2005 Aug;125(8):870-6. doi: 10.1080/00016480510031489. Acta Otolaryngol. 2005. PMID: 16158535 Review.
-
[Cochlear implantation--the Viennese surgical technique].Wien Klin Wochenschr. 2000 Jun 2;112(11):473-6. Wien Klin Wochenschr. 2000. PMID: 10890123 Review. German.
Cited by
-
Effect of Electrode Insertion Angle on Cochlear Implantation Outcomes in Adult and Children Patients with Sensorineural Hearing Loss.Oxid Med Cell Longev. 2022 Aug 23;2022:9914716. doi: 10.1155/2022/9914716. eCollection 2022. Oxid Med Cell Longev. 2022. PMID: 36052159 Free PMC article.
-
Evaluation of the Relationship between the NRT-Ratio, Cochlear Anatomy, and Insertions Depth of Perimodiolar Cochlear Implant Electrodes.Biomed Res Int. 2015;2015:706253. doi: 10.1155/2015/706253. Epub 2015 Dec 29. Biomed Res Int. 2015. PMID: 26839885 Free PMC article.
-
Comparative Analysis of Robotics-Assisted and Manual Insertions of Cochlear Implant Electrode Arrays.Otol Neurotol. 2022 Dec 1;43(10):1155-1161. doi: 10.1097/MAO.0000000000003707. Epub 2022 Oct 6. Otol Neurotol. 2022. PMID: 36201552 Free PMC article.
-
Electrophysiological detection of electrode fold-over in perimodiolar cochlear implant electrode arrays: a multi-center study case series.Eur Arch Otorhinolaryngol. 2020 Jan;277(1):31-35. doi: 10.1007/s00405-019-05653-9. Epub 2019 Sep 24. Eur Arch Otorhinolaryngol. 2020. PMID: 31552525
-
Effect on vestibular function of cochlear implantation by partial deafness treatment-electro acoustic stimulation (PDT-EAS).Eur Arch Otorhinolaryngol. 2019 Jul;276(7):1951-1959. doi: 10.1007/s00405-019-05425-5. Epub 2019 May 3. Eur Arch Otorhinolaryngol. 2019. PMID: 31053967 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical