Simultaneous determination of acetylsalicylic acid and caffeine in pharmaceutical formulation by first derivative synchronous fluorimetric method

Affiliations

Simultaneous determination of acetylsalicylic acid and caffeine in pharmaceutical formulation by first derivative synchronous fluorimetric method

Mohammad Mainul Karim et al. J Fluoresc. 2006 Sep.

Abstract

A sensitive, rapid, and specific assay has been developed for the simultaneous determination of acetylsalicylic acid and caffeine in commercial tablets based on their natural fluorescence. The mixture of these drugs was resolved by first derivative synchronous fluorimetric technique using two scans. At Deltalambda=106 nm, using first derivative synchronous scanning, only acetylsalicylic acid yields a detectable signal at 316 nm (peak to zero method) which is unaffected by caffeine. At Deltalambda=30 nm, the signal of caffeine at 288 nm (peak to zero method) is not affected by acetylsalicylic acid. The range of application is between 0.021 and 41.62 microg ml(-1) (correlation coefficient, R=0.9995) for acetylsalicylic acid and between 0.4486 and 44.86 microg ml(-1) (correlation coefficient, R=0.99786) for caffeine. The recovery range of 98.40-102% for acetylsalicylic acid and 90-100.5% for caffeine from their synthetic mixture was reported. Overall recovery of both compounds about 97-99% for acetylsalicylic acid and 97-98% for caffeine was obtained from real sample analysis. The detection limits are 0.0013 microg ml(-1) and 0.0306 microg ml(-1) for acetylsalicylic acid and caffeine, respectively. The relative standard deviation (n=10) for 20 microg ml(-1) of acetylsalicylic acid is 2.75% and for 2.2 microg ml(-1)of caffeine is 1.7%.

PubMed Disclaimer

References

    1. J Chromatogr Sci. 2004 Mar;42(3):155-60 - PubMed
    1. J Sep Sci. 2004 May;27(7-8):529-36 - PubMed
    1. Talanta. 1998 Mar;45(5):969-76 - PubMed
    1. Anal Chem. 1978 Mar;50(3):396-401 - PubMed
    1. Electrophoresis. 2004 Feb;25(4-5):615-21 - PubMed

Publication types

MeSH terms

LinkOut - more resources