Lack of sphygmomanometer calibration causes over- and under-detection of hypertension: a computer simulation study
- PMID: 16957551
- DOI: 10.1097/01.hjh.0000244940.11675.82
Lack of sphygmomanometer calibration causes over- and under-detection of hypertension: a computer simulation study
Abstract
Objective: To estimate the contribution of inadequate sphygmomanometer calibration to over- and under-detection of hypertension.
Design: Monte Carlo simulation of the measurement of blood pressure (BP) of a population with calibrated and uncalibrated sphygmomanometers. Simulated BP measurements included systematic sphygmomanometer error and random variability.
Main outcome measures: The percentage of hypertensive adults (BP > 140/90 mmHg) not detected and the percentage of adults incorrectly classified hypertensive due to sphygmomanometer error. The percentage of the false positives and false negatives attributable to sphygmomanometer error. The number of additional visits patients need to make to obtain the same improvement in hypertension detection as is obtained by sphygmomanometer calibration.
Results: After three visits, uncalibrated sphygmomanometer error causes 20 and 28% of all undetected adult systolic and diastolic hypertension, respectively, and 15 and 31% of all falsely detected adult systolic and diastolic hypertension. In some groups, under-detection is worse; for example, sphygmomanometer error causes 27% of all missed systolic hypertension in 35-44-year-old females. In some age groups, over-detection is worse; for example, after three visits, sphygmomanometer error causes 63 and 50% of falsely detected systolic and diastolic hypertension in 18-24-year-old females, respectively. In-service sphygmomanometer calibration achieves the same or greater improvement in hypertension detection as an additional two visits.
Conclusions: Uncalibrated sphygmomanometers are a preventable cause of clinically significant over- and under-detection of hypertension. Sphygmomanometers should be calibrated regularly by accredited organizations or technicians. Standards and guidelines governing sphygmomanometers in service should be revised. Sphygmomanometer calibration is a cost-effective way of improving hypertension detection.
Comment in
-
Accuracy of blood pressure measurement: sphygmomanometer calibration and beyond.J Hypertens. 2006 Oct;24(10):1915-8. doi: 10.1097/01.hjh.0000244935.19299.f5. J Hypertens. 2006. PMID: 16957546 No abstract available.
Similar articles
-
Lax sphygmomanometer standard causes overdetection and underdetection of hypertension: a computer simulation study.Blood Press Monit. 2008 Apr;13(2):91-9. doi: 10.1097/MBP.0b013e3282f7691c. Blood Press Monit. 2008. PMID: 18347443
-
Sphygmomanometer calibration--why, how and how often?Aust Fam Physician. 2007 Oct;36(10):834-8. Aust Fam Physician. 2007. PMID: 17925905
-
The extent and implications of sphygmomanometer calibration error in primary care.J Hum Hypertens. 2001 Sep;15(9):587-91. doi: 10.1038/sj.jhh.1001241. J Hum Hypertens. 2001. PMID: 11550103
-
Ten-year blood pressure trends in nonhypertensive inhabitants of La Plata, Argentina.Can J Cardiol. 1998 Jul;14(7):917-22. Can J Cardiol. 1998. PMID: 9706276 Review.
-
Alternatives to the mercury sphygmomanometer.J Public Health Policy. 2011 Feb;32(1):107-20. doi: 10.1057/jphp.2010.38. Epub 2010 Nov 25. J Public Health Policy. 2011. PMID: 21109765 Review.
Cited by
-
Predicted impact of various clinical practice strategies on cardiovascular risk for the treatment of hypertension: a clinical trial simulation study.J Pharmacokinet Pharmacodyn. 2014 Dec;41(6):693-704. doi: 10.1007/s10928-014-9394-9. Epub 2014 Oct 18. J Pharmacokinet Pharmacodyn. 2014. PMID: 25326066 Clinical Trial.
-
Type and accuracy of sphygmomanometers in primary care: a cross-sectional observational study.Br J Gen Pract. 2011 Sep;61(590):e598-603. doi: 10.3399/bjgp11X593884. Br J Gen Pract. 2011. PMID: 22152749 Free PMC article.
-
Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone.Sensors (Basel). 2023 Oct 9;23(19):8340. doi: 10.3390/s23198340. Sensors (Basel). 2023. PMID: 37837170 Free PMC article.
-
Agreement of blood pressure measurements between random-zero and standard mercury sphygmomanometers.Am J Med Sci. 2008 Nov;336(5):373-8. doi: 10.1097/MAJ.0b013e31816956ac. Am J Med Sci. 2008. PMID: 19011391 Free PMC article.
-
Results of a project to calibrate mercury sphygmomanometer blood pressure-measuring devices in Egypt.J Hum Hypertens. 2021 Oct;35(10):921-926. doi: 10.1038/s41371-020-00424-0. Epub 2020 Oct 9. J Hum Hypertens. 2021. PMID: 33037321 Free PMC article.