Measurement of intraocular pressure after LASIK by dynamic contour tonometry
- PMID: 16961298
- DOI: 10.1007/BF02829580
Measurement of intraocular pressure after LASIK by dynamic contour tonometry
Abstract
Changes of corneal properties induced by laser in situ keratomileusis (LASIK) results in low inaccurate intraocular pressure (IOP) readings by Goldmann applanation tonometry (GAT). Before and after LASIK, the applied value of IOP, measured by dynamic contour tonometry (DCT) in comparison to GAT, was evaluated. Before and 1, 4 weeks after LASIK, the IOP in 30 cases (60 eyes) was measured by GAT and DCT respectively. The obtained results were statistically processed by SPSS11.5 statistical software. The results showed that central corneal thickness (CCT) could affect GAT measurements but not DCT measurements. The comparison of IOP one and 4 weeks after LASIK revealed that the readings from GAT was separately decreased by 5.00 +/- 1.12 and 5.45 +/- 1.13 mmHg as compared with those before LASIK, while those from DCT had no significant difference. It was concluded that LASIK-induced changes of CCT could influence the accuracy of GAT measurements, but had no influence on those from DCT. DCT was more beneficial to the measurements of IOP in normal eyes and those subject to LASIK surgery.
Similar articles
-
Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis.Invest Ophthalmol Vis Sci. 2003 Sep;44(9):3790-4. doi: 10.1167/iovs.02-0946. Invest Ophthalmol Vis Sci. 2003. PMID: 12939293 Clinical Trial.
-
Effect of Laser in Situ Keratomileusis on Schiøtz, Goldmann, and Dynamic Contour Tonometric Measurements.J Glaucoma. 2016 Apr;25(4):e419-23. doi: 10.1097/IJG.0000000000000338. J Glaucoma. 2016. PMID: 26550981
-
Effectiveness of the Goldmann Applanation Tonometer, the Dynamic Contour Tonometer, the Ocular Response Analyzer and the Corvis ST in Measuring Intraocular Pressure following FS-LASIK.Curr Eye Res. 2020 Feb;45(2):144-152. doi: 10.1080/02713683.2019.1660794. Epub 2019 Dec 26. Curr Eye Res. 2020. PMID: 31869261
-
The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis.Indian J Ophthalmol. 2024 May 1;72(Suppl 3):S398-S403. doi: 10.4103/IJO.IJO_907_23. Epub 2023 Dec 15. Indian J Ophthalmol. 2024. PMID: 38099358 Free PMC article. Review.
-
Interface fluid after laser in situ keratomileusis.J Refract Surg. 2003 Jul-Aug;19(4):455-9. doi: 10.3928/1081-597X-20030701-13. J Refract Surg. 2003. PMID: 12899478 Review.