Prospective comparison of high-dose plateletpheresis with the latest apheresis systems on the same donors
- PMID: 16965590
- DOI: 10.1111/j.1537-2995.2006.00928.x
Prospective comparison of high-dose plateletpheresis with the latest apheresis systems on the same donors
Abstract
Background: To improve productivity of automated platelet (PLT) collection, the industry has introduced new instruments or modifications to existing equipment.
Study design and methods: With the same 8 donors for double (DDC) and triple-dose PLT collection (TDC), the Baxter Amicus (AM), the Haemonetics MCS Plus (MCS+), and the Gambro Trima Accel (TA) were evaluated focusing on yield, duration, and citrate donor load. Target endpoints were set at 5.5 x 10(11) to 6.0 x 10(11) PLTs (DDC) and 7.5 x 10(11) to 8.0 x 10(11) PLTs (TDC) in up to 100 and 120 minutes' donation time, respectively.
Results: TA was the most efficient system (74.5 +/- 3.9%) with significant differences from AM (71.1 +/- 3.9%; p = 0.028) and MCS+ (64.0 +/- 7.7%; p = 0.002). TA had advantages over AM for collection rate (10.9 x 10(9) +/- 2.2 x 10(9) vs. 10.1 x 10(9) +/- 1.5 x 10(9) PLTs/min; p = 0.382), whole blood processed (3928 +/- 611 mL vs. 4219 +/- 727 mL; p = 0.382), and time to obtain an established standard dose (TSD 2.5(EU), 30.2 +/- 5.6 vs. 37.7 +/- 5.5 min; TSD 3.5(US), 42.2 +/- 7.8 min vs. 52.7 +/- 7.7 min; p = 0.015), whereas AM was slightly superior in PLT yield (2.81 x 10(11) +/- 0.21 x 10(11) vs. 2.76 x 10(11) +/- 0.31 x 10(11)/unit; p = 0.645). Owing to the lowest draw (42.3 +/- 3.2 mL/min; p < 0.001) and collection rates (6.0 x 10(11) +/- 1.5 x 10(11)/min; p = 0.021), MCS+ was the slowest significantly (p < 0.001) but compensated with fewer citrate reactions owing to lower citrate infusion rates (0.78 +/- 0.11 mL/min/L; p = 0.028).
Conclusion: High-dose plateletpheresis was performed efficiently and safely with all three instruments. AM had advantages in PLT yield, and MCS+, in donor comfort. TA was the fastest in obtaining an established standard dose and, because of this advantage, the machine with the highest practical impact in routine use.
Similar articles
-
Evaluation of concurrent collection of in-line filtered platelets and packed red blood cells by multicomponent apheresis with three last-generation apparatuses.Vox Sang. 2006 Jul;91(1):47-55. doi: 10.1111/j.1423-0410.2006.00774.x. Vox Sang. 2006. PMID: 16756601
-
Paired comparison of Gambro Trima Accel versus Baxter Amicus single-needle plateletpheresis.Transfusion. 2004 Nov;44(11):1612-20. doi: 10.1111/j.0041-1132.2004.04129.x. Transfusion. 2004. PMID: 15504167
-
Frequently used plateletpheresis techniques result in variable target yields and platelet recruitment of donors.Transfusion. 2005 May;45(5):788-97. doi: 10.1111/j.1537-2995.2005.04353.x. Transfusion. 2005. PMID: 15847670
-
Blood component collection by apheresis.J Clin Apher. 2006 Jul;21(2):142-51. doi: 10.1002/jca.20043. J Clin Apher. 2006. PMID: 15880369 Review.
-
Safety of donating multiple products in a single apheresis collection: are we expecting too much?J Clin Apher. 2003;18(3):135-40. doi: 10.1002/jca.10069. J Clin Apher. 2003. PMID: 14569607 Review.
Cited by
-
A Study on Influence of Donor Hematocrit on the Procedural Parameters of Concentrated Single Donor Platelets Collected by Two Apheresis Devices.Indian J Hematol Blood Transfus. 2020 Jan;36(1):135-140. doi: 10.1007/s12288-019-01163-0. Epub 2019 Jul 27. Indian J Hematol Blood Transfus. 2020. PMID: 32158096 Free PMC article.
-
Monitoring of platelet activation in platelet concentrates using transmission electron microscopy.Transfus Med Hemother. 2013 Apr;40(2):101-7. doi: 10.1159/000350034. Epub 2013 Mar 14. Transfus Med Hemother. 2013. PMID: 23652838 Free PMC article.
-
Analysis of Donor Safety in High Yield Plateletpheresis Procedures: An Experience from Tertiary Care Hospital in South India.Indian J Hematol Blood Transfus. 2020 Jul;36(3):542-549. doi: 10.1007/s12288-020-01266-z. Epub 2020 Feb 27. Indian J Hematol Blood Transfus. 2020. PMID: 32647430 Free PMC article.
-
Single-donor platelet apheresis: observational comparison of the new Haemonetics Universal Platelet protocol with the previous Concentrated Single Donor Platelet protocol.Blood Transfus. 2014 Apr;12(2):220-5. doi: 10.2450/2013.0119-13. Epub 2013 Dec 4. Blood Transfus. 2014. PMID: 24333087 Free PMC article.
-
Immature Platelet Count Levels as a Novel Quality Marker in Plateletpheresis.Indian J Hematol Blood Transfus. 2018 Oct;34(4):684-690. doi: 10.1007/s12288-018-0965-6. Epub 2018 May 14. Indian J Hematol Blood Transfus. 2018. PMID: 30369741 Free PMC article.
Publication types