Effect of 3D ultrasound probes on the accuracy of electromagnetic tracking systems

Affiliations

Effect of 3D ultrasound probes on the accuracy of electromagnetic tracking systems

Mark Hastenteufel et al. Ultrasound Med Biol. 2006 Sep.

Abstract

In the last few years, 3D ultrasound probes have became readily available. New fields of image-guided surgery applications are opened by attaching small electromagnetic position sensors to 3D ultrasound probes. However, nothing is known about the distortions caused by 3D ultrasound probes regarding electromagnetic sensors. Several trials were performed to investigate error-proneness of state-of-the-art electromagnetic tracking systems when used in combination with 3D ultrasound probes. It was found that 3D ultrasound probes do distort electromagnetic sensors more than 2D probes do. When attaching electromagnetic sensors to 3D probes, maximum errors of 5 mm up to 119 mm occur. The distortion strongly depends on the electromagnetic technology as well on the probe technology used. Thus, for 3D ultrasound-guided applications using electromagnetic tracking technology, the interference of ultrasound probes and electromagnetic sensors have to be checked carefully.

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources