Transcutaneous monitoring of partial pressure of carbon dioxide in the elderly patient: a prospective, clinical comparison with end-tidal monitoring
- PMID: 16980160
- DOI: 10.1016/j.jclinane.2006.02.007
Transcutaneous monitoring of partial pressure of carbon dioxide in the elderly patient: a prospective, clinical comparison with end-tidal monitoring
Erratum in
- J Clin Anesth. 2006 Nov;18(7):559
Abstract
Study objective: To evaluate the accuracy and precision of estimation of partial pressure of carbon dioxide (Pa(CO2)) using end-tidal or transcutaneous CO2 (TcP(CO2)) measurements during mechanical ventilation in the elderly patient.
Design: A prospective, observational study was conducted.
Settings: The study was done in the anesthesia department of a university hospital.
Patients: Seventeen anesthetized, mechanically ventilated patients older than 60 years were studied.
Interventions and measurements: During standard sevoflurane anesthesia, and after proper calibration and an equilibration time of 30 minutes with stable hemodynamic and respiratory variables, arterial (Pa(CO2)), end-tidal (Pet(CO2)), and transcutaneous (TcP(CO2)) CO2 partial pressures were determined. In each patient, 1 to 5 sample sets (Pa(CO2), Pet(CO2), and TcP(CO2)) were obtained.
Main results: A total of 45 sample sets were obtained from the patients studied. The Pa(CO2) values ranged between 21 and 58 mm Hg. The Pa(CO2) - Pet(CO2) tension gradient was 6 +/- 5 mmHg (95% confidence interval, -3 to 16 mmHg), whereas the Pa(CO2) - TcP(CO2) tension gradient was 2 +/- 4 mmHg (95% confidence interval, -6 to 9 mmHg) (P = 0.0005). The absolute value of the difference between Pa(CO2) and Pet(CO2) was 3 mm Hg or less in 7 of 45 sample sets (15%), whereas the absolute value of the difference between Pa(CO2) and TcP(CO2) was 3 mm Hg or less in 21 of 45 sample sets (46%) (P = 0.003). Linear regression analysis for TcP(CO2) versus Pa(CO2) showed a slope of 0.84 (r(2) = 0.73), whereas the linear regression analysis for Pet(CO2) versus Pa(CO2) showed a slope of 0.54 (r(2) = 0.50).
Conclusion: Transcutaneous monitoring of CO(2) partial pressure gives a more accurate estimation of arterial CO(2) partial pressure than does Pet(CO2) monitoring.
Similar articles
-
[Simultaneous measurements of end-expiratory and transcutaneous carbon dioxide partial pressure in ventilated premature and newborn infants].Klin Padiatr. 1997 Mar-Apr;209(2):47-53. doi: 10.1055/s-2008-1043927. Klin Padiatr. 1997. PMID: 9198671 German.
-
End tidal carbon dioxide monitoring in spontaneously breathing, nonintubated patients. A clinical comparison between conventional sidestream and microstream capnometers.Minerva Anestesiol. 2001 Apr;67(4):161-4. Minerva Anestesiol. 2001. PMID: 11376503 Clinical Trial.
-
An evaluation of capnography monitoring during the apnoea test in brain-dead patients.Eur J Anaesthesiol. 2007 Oct;24(10):868-75. doi: 10.1017/S0265021507000725. Epub 2007 Jun 20. Eur J Anaesthesiol. 2007. PMID: 17579950 Clinical Trial.
-
Transcutaneous carbon dioxide monitoring in infants and children.Paediatr Anaesth. 2009 May;19(5):434-44. doi: 10.1111/j.1460-9592.2009.02930.x. Epub 2009 Feb 19. Paediatr Anaesth. 2009. PMID: 19236597 Review.
-
Ventilation monitoring during monitored anesthesia care: a review.AANA J. 1994 Dec;62(6):521-6. AANA J. 1994. PMID: 7879583 Review.
Cited by
-
Noninvasive carbon dioxide monitoring in pediatric patients undergoing laparoscopic surgery: transcutaneous vs. end-tidal techniques.BMC Pediatr. 2023 Jan 14;23(1):20. doi: 10.1186/s12887-023-03836-2. BMC Pediatr. 2023. PMID: 36639787 Free PMC article.
-
The use of transcutaneous CO2 monitoring in cardiac arrest patients: a feasibility study.Scand J Trauma Resusc Emerg Med. 2014 Nov 29;22:70. doi: 10.1186/s13049-014-0070-2. Scand J Trauma Resusc Emerg Med. 2014. PMID: 25471936 Free PMC article.
-
Cerebral vasoreactivity during hypercapnia is reset by augmented sympathetic influence.J Appl Physiol (1985). 2011 Feb;110(2):352-8. doi: 10.1152/japplphysiol.00802.2010. Epub 2010 Nov 11. J Appl Physiol (1985). 2011. PMID: 21071587 Free PMC article.
-
Transcutaneous PCO2 monitors are more accurate than end-tidal PCO2 monitors.J Anesth. 2009;23(2):198-202. doi: 10.1007/s00540-008-0734-z. Epub 2009 May 15. J Anesth. 2009. PMID: 19444557
-
Correlation of end-tidal carbon dioxide with arterial carbon dioxide in mechanically ventilated patients.Arch Trauma Res. 2012 Summer;1(2):58-62. doi: 10.5812/atr.6444. Epub 2012 Aug 21. Arch Trauma Res. 2012. PMID: 24396744 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources