[Medium-term results of the cementless femoral neck prosthesis CUT]
- PMID: 16991063
- DOI: 10.1055/s-2006-942265
[Medium-term results of the cementless femoral neck prosthesis CUT]
Abstract
Aim: The present work aims at evaluating the clinical and radiological midterm results of the femoral neck prosthesis CUT.
Method: 67 femoral neck prostheses of the type CUT were implanted in 63 patients between 4/1999 and 5/2003 (average age 56 years). At an average follow-up time of 5.1 years (min.: 2.9; max.: 6.9) 53 patients with 57 CUT prostheses were examined clinically with the Harris hip score and radiologically. All together we determined the state of 64 CUT prostheses (96%).
Results: The mean Harris hip score improved from 55 points preoperatively to 90.4 points at the last follow-up and is assessed as a good result. A total of 7 CUT prostheses (11%) had been revised. 3 CUT prostheses (4.7%) had been revised because of aseptic loosening, 2 (3.1%) because of persisting postoperative thigh pain, 1 (1.6%) because of vertical migration of an unsuitable CUT prostheses a few days after the operation and 1 (1.6%) because of septic loosening. The survival rate according to Kaplan-Meier was 89.1% at 5.1 years. If the femoral neck was resected too widely there was an increased rate of horizontal migration with aseptic loosening of the CUT prostheses.
Conclusion: In the midterm follow-up the CUT prosthesis shows a higher loosening rate as compared with cementless standard stems.
Similar articles
-
Cementless CUT femoral neck prosthesis: increased rate of aseptic loosening after 5 years.Acta Orthop. 2007 Oct;78(5):616-21. doi: 10.1080/17453670710014301. Acta Orthop. 2007. PMID: 17966020
-
[Cementless femoral neck prosthesis CUT--clinical and radiological results after 5 years].Z Orthop Unfall. 2010 Aug;148(4):413-9. doi: 10.1055/s-0030-1250150. Epub 2010 Aug 16. Z Orthop Unfall. 2010. PMID: 20714982 German.
-
[Long-term results of the Schenker cementless femoral component].Acta Chir Orthop Traumatol Cech. 2004;71(2):93-100. Acta Chir Orthop Traumatol Cech. 2004. PMID: 15151096 Czech.
-
Design aspects and clinical performance of the thrust plate hip prosthesis.Proc Inst Mech Eng H. 2007 Jan;221(1):29-37. doi: 10.1243/09544119JEIM111. Proc Inst Mech Eng H. 2007. PMID: 17315766 Review.
-
Results of alloarthroplasty of the hip joint.Arch Orthop Trauma Surg (1978). 1980;96(4):247-58. doi: 10.1007/BF00432862. Arch Orthop Trauma Surg (1978). 1980. PMID: 6994678 Review.
Cited by
-
Maintenance of bone mineral density after implantation of a femoral neck hip prosthesis.BMC Musculoskelet Disord. 2008 Jan 31;9:17. doi: 10.1186/1471-2474-9-17. BMC Musculoskelet Disord. 2008. PMID: 18237437 Free PMC article.
-
Uncemented short stems in primary total hip arthroplasty: The state of the art.EFORT Open Rev. 2018 May 9;3(5):149-159. doi: 10.1302/2058-5241.3.170052. eCollection 2018 May. EFORT Open Rev. 2018. PMID: 29951251 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources