The effects of apparatus dead space on P(aCO2) in patients receiving lung-protective ventilation
- PMID: 17005059
The effects of apparatus dead space on P(aCO2) in patients receiving lung-protective ventilation
Abstract
Background: Lung-protective ventilation using tidal volume (V(T)) of 4-6 mL/kg (predicted body weight) reduces mortality (compared with traditional V(T)) in patients with acute respiratory distress syndrome and acute lung injury. Standardized use of lower V(T) can result in respiratory acidosis and has raised new concerns about the appropriate configuration of the ventilator circuit, especially in regard to the dead space (V(D)) of the apparatus. We hypothesized that, with a patient receiving lung-protective ventilation, the removal of all apparatus dead space from the circuit would reduce P(aCO2) and allow a reduction in minute ventilation.
Methods: All the studied patients met the American-European consensus-conference criteria for acute respiratory distress syndrome/acute lung injury, were receiving a lung-protective ventilation strategy, were > 18 years of age, and were hemodynamically stable. We prospectively tested 3 different ventilator-circuit configurations, in random sequence, for 15 min each: (1) standard hygroscopic heat-and-moisture exchanger (HME) with 15-cm flexible tubing, (2) 15-cm flexible tubing only, (3) no HME or flexible tubing. V(T), respiratory rate, positive end-expiratory pressure, and fraction of inspired oxygen were maintained constant throughout the study, and exhaled CO2 was measured continuously. Physiologic dead space (V(D)/V(T)) was calculated using the Enghoff modification of the Bohr equation.
Results: Seven patients were studied. Removal of the HME from the circuit significantly decreased V(D)/V(T) (by approximately 6%) and P(aCO2) (by approximately 5 mm Hg). Removal of both the HME and flexible tubing from the circuit reduced V(D)/V(T) by an additional 5%, and P(aCO2) by an additional 6 mm Hg. With both circuit-configuration changes, minute ventilation fell from a mean of 11.51 L/min to 10.35 L/min, and pH increased from 7.30 to 7.38. Carbon-dioxide production did not change significantly.
Conclusion: In patients receiving lower-V(T) ventilation, removing all the apparatus V(D) from the ventilator circuit reduces P(aCO2) and increases pH, at a lower minute ventilation. This information will help guide ventilator-circuit configuration for patients receiving lung-protective ventilation.
Similar articles
-
The effects of passive humidifier dead space on respiratory variables in paralyzed and spontaneously breathing patients.Respir Care. 2000 Mar;45(3):306-12. Respir Care. 2000. PMID: 10771799 Clinical Trial.
-
Work of breathing during lung-protective ventilation in patients with acute lung injury and acute respiratory distress syndrome: a comparison between volume and pressure-regulated breathing modes.Respir Care. 2005 Dec;50(12):1623-31. Respir Care. 2005. PMID: 16318643 Clinical Trial.
-
Setting the frequency-tidal volume pattern.Respir Care. 2002 Mar;47(3):266-74; discussion 274-8. Respir Care. 2002. PMID: 11874606 Review.
-
Implementation of a low tidal volume ventilation protocol for patients with acute lung injury or acute respiratory distress syndrome.Respir Care. 2001 Oct;46(10):1024-37. Respir Care. 2001. PMID: 11572755 Review.
-
Effects of continuous tracheal gas insufflation during pressure limited ventilation on pulmonary surfactant in rabbits with acute lung injury.Chin Med J (Engl). 2006 Sep 5;119(17):1415-20. Chin Med J (Engl). 2006. PMID: 16989741
Cited by
-
Humidification and heating of inhaled gas in patients with artificial airway. A narrative review.Rev Bras Ter Intensiva. 2018 Mar;30(1):86-97. doi: 10.5935/0103-507x.20180015. Rev Bras Ter Intensiva. 2018. PMID: 29742220 Free PMC article. Review.
-
The effect of a pediatric heat and moisture exchanger on dead space in healthy pediatric anesthesia.Korean J Anesthesiol. 2012 May;62(5):418-22. doi: 10.4097/kjae.2012.62.5.418. Epub 2012 May 24. Korean J Anesthesiol. 2012. PMID: 22679537 Free PMC article.
-
Removal of a catheter mount and heat-and-moisture exchanger improves hypercapnia in patients with acute respiratory distress syndrome: A retrospective observational study.Medicine (Baltimore). 2021 Sep 10;100(36):e27199. doi: 10.1097/MD.0000000000027199. Medicine (Baltimore). 2021. PMID: 34516524 Free PMC article.
-
Tidal volume and plateau pressure use for acute lung injury from 2000 to present: a systematic literature review.Crit Care Med. 2014 Oct;42(10):2278-89. doi: 10.1097/CCM.0000000000000504. Crit Care Med. 2014. PMID: 25098333 Free PMC article.
-
Low Tidal Volume Ventilation in Patients without Acute Respiratory Distress Syndrome: A Paradigm Shift in Mechanical Ventilation.Crit Care Res Pract. 2012;2012:416862. doi: 10.1155/2012/416862. Epub 2012 Mar 27. Crit Care Res Pract. 2012. PMID: 22536499 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources
Miscellaneous