. 2002 Jul 1;2(3):74-8.

Electromagnetic interference on pacemakers

Affiliations

Electromagnetic interference on pacemakers

Okan Erdogan. Indian Pacing Electrophysiol J. .

Abstract

External sources, either within or outside the hospital environment, may interfere with the appropriate function of pacemakers which are being implanted all around the world in current medical practice. The patient and the physician who is responsible for follow-up of the pacing systems may be confronted with some specific problems regarding the various types of electromagnetic interference (EMI). To avoid these unwanted EMI effects one must be aware of this potential problem and need to take some precautions. The effects of EMI on pacemaker function and precautions to overcome some specific problems were discussed in this review article. There are many sources of EMI interacting with pacemakers. Magnetic resonance imaging creates real problem and should be avoided in pacemaker patients. Cellular phones might be responsible for EMI when they were held on the same side with the pacemaker. Otherwise they don't cause any specific type of interaction with pacemakers. Sale security systems are not a problem if one walks through it without lingering in or near it. Patients having unipolar pacemaker systems are prone to develop EMI because of pectoral muscle artifacts during vigorous active physical exercise.

PubMed Disclaimer

References

    1. Das G, Eaton J. Pacemaker malfunction following transthoracic countershock. PACE. 1981;4:487–490. - PubMed
    1. Altamura G, Bianconi L, Lo Bianco F, et al. Transthoracic DC shock may represent a serious hazard in pacemaker dependent patients. PACE. 1995;18:194–198. - PubMed
    1. Gould L, Patel S, Gomez GL, et al. Pacemaker failure following external defibrillation. PACE. 1981;4:575, 577. - PubMed
    1. Proclemer A, Facchin D, Pagnutti C, et al. Safety of pacemaker implantation prior to radiofrequency ablation of atrioventricular junction in a single session procedure. PACE. 2000;23:998, 1002. - PubMed
    1. Pfeiffer B, Tebbenjohanns J, Schumacher B, et al. Pacemaker function during radiofrequency catheter ablation. PACE. 1995;18:1037–1044. - PubMed

LinkOut - more resources