Feasibility and safety of a novel technology for pacing without leads
- PMID: 17018352
- DOI: 10.1016/j.hrthm.2006.06.012
Feasibility and safety of a novel technology for pacing without leads
Abstract
Background: Pacemaker lead complications and failures remain clinical problems. New devices incorporating three leads are associated with even greater limitations.
Objectives: The purpose of this study was to investigate the feasibility and safety of a technology enabling cardiac pacing without leads in an acute porcine model.
Methods: The system is composed of an ultrasound transmitter delivering energy from the chest wall to a receiver-electrode in contact with the myocardium that then converts the ultrasound energy to electrical energy sufficient to pace. In five feasibility studies, the receiver-electrodes were attached to the tip of a catheter to facilitate intracardiac positioning at pacing sites. In six safety studies (five treatment and one sham), ultrasound energy was transmitted to both chest walls, and histopathologic examinations were performed to evaluate bioeffects due to ultrasound energy transmission.
Results: In five feasibility studies, direct and ultrasound-mediated electrical pacing was demonstrated at 30 sites in the right atrium, right ventricle, and left ventricle, at direct electrical pacing outputs of 1.4 +/- 0.6 V and ultrasound-mediated electrical pacing outputs of 1.8 +/- 0.9 V. The mechanical index was 0.6 +/- 0.4 at the receiver site during ultrasound-mediated pacing at a depth of 11.2 +/- 2.4 cm from the chest wall. Using two receiver-electrode catheters, biventricular pacing was demonstrated in all studies. In five safety study treatment animals at a similar depth, the peak mechanical index was 2.3, and the thermal index was 0.4. Microscopic evaluation revealed no evidence of mechanical or thermal bioeffects.
Conclusion: The feasibility and safety of this novel technology for pacing without leads has been demonstrated acutely in animals.
Comment in
-
In a footnote, at least.Heart Rhythm. 2006 Oct;3(10):1207-9. doi: 10.1016/j.hrthm.2006.06.025. Epub 2006 Jul 8. Heart Rhythm. 2006. PMID: 17018353 No abstract available.
Similar articles
-
First human demonstration of cardiac stimulation with transcutaneous ultrasound energy delivery: implications for wireless pacing with implantable devices.J Am Coll Cardiol. 2007 Aug 28;50(9):877-83. doi: 10.1016/j.jacc.2007.04.081. Epub 2007 Aug 13. J Am Coll Cardiol. 2007. PMID: 17719475
-
Temporary leadless pacing in heart failure patients with ultrasound-mediated stimulation energy and effects on the acoustic window.Heart Rhythm. 2009 Jun;6(6):742-8. doi: 10.1016/j.hrthm.2009.02.025. Epub 2009 Feb 24. Heart Rhythm. 2009. PMID: 19427274
-
First-in-man implantation of leadless ultrasound-based cardiac stimulation pacing system: novel endocardial left ventricular resynchronization therapy in heart failure patients.Europace. 2013 Aug;15(8):1191-7. doi: 10.1093/europace/eut124. Epub 2013 May 23. Europace. 2013. PMID: 23703364
-
Leadless Cardiac Pacemakers: Pacing Paradigm Change.Curr Cardiol Rep. 2015 Aug;17(8):68. doi: 10.1007/s11886-015-0619-3. Curr Cardiol Rep. 2015. PMID: 26233700 Review.
-
Multi-Axes Lead With Tetrahedral Electrode Tip for Cardiac-Implantable Devices: Creative Concept for Pacing and Sensing Technology.Can J Cardiol. 2021 Nov;37(11):1808-1817. doi: 10.1016/j.cjca.2021.07.013. Epub 2021 Jul 29. Can J Cardiol. 2021. PMID: 34333028 Review.
Cited by
-
Endocardial pacing: the wave of the future?Curr Cardiol Rep. 2012 Oct;14(5):547-51. doi: 10.1007/s11886-012-0298-2. Curr Cardiol Rep. 2012. PMID: 22825920 Review.
-
Wireless microstimulators for neural prosthetics.Crit Rev Biomed Eng. 2011;39(1):63-77. doi: 10.1615/critrevbiomedeng.v39.i1.50. Crit Rev Biomed Eng. 2011. PMID: 21488815 Free PMC article. Review.
-
The Negative Chronotropic Effect in Rat Heart Stimulated by Ultrasonic Pulses: Role of Sex and Age.J Ultrasound Med. 2017 Apr;36(4):799-808. doi: 10.7863/ultra.16.02017. Epub 2017 Jan 10. J Ultrasound Med. 2017. PMID: 28072471 Free PMC article.
-
Chronic ventricular pacing in children: toward prevention of pacing-induced heart disease.Heart Fail Rev. 2011 May;16(3):305-14. doi: 10.1007/s10741-010-9207-1. Heart Fail Rev. 2011. PMID: 21107685 Free PMC article. Review.
-
Wireless communication with implanted medical devices using the conductive properties of the body.Expert Rev Med Devices. 2011 Jul;8(4):427-33. doi: 10.1586/erd.11.16. Expert Rev Med Devices. 2011. PMID: 21728728 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical