. 2007 Apr;17(4):1118-24.
doi: 10.1007/s00330-006-0421-0. Epub 2006 Oct 10.

MR-compatible assistance system for punction in a high-field system: device and feasibility of transgluteal biopsies of the prostate gland

Affiliations

MR-compatible assistance system for punction in a high-field system: device and feasibility of transgluteal biopsies of the prostate gland

Stephan Zangos et al. Eur Radiol. 2007 Apr.

Abstract

We present the first cadavic study results concerning the feasibility of the use of an MR-guided assistance system, Innomotion (Innomedic, Herxheim, Germany), for accurate and consistent placement of percutaneous needles in the prostate gland. The MR-compatible assistance system consists of a C-arch, guiding arm and application module (AMO). T1-weighted fast low angle shot (FLASH) 2-D-GRE sequence (TR/TE=110/4 ms) and T2-weighted turbo spin-echo (TSE)-sequences (TR/TE=3200/97 ms) in transversal orientation were used for the monitoring of the punction of the prostate gland. Planning and control of the intervention is to be made outside the scanner room on a desktop computer that receives DICOM images from the scanner. Servopneumatic drives move the AMO to the insertion point. The physician has to introduce the punction needle manually. The mean deviation of the needle tip to the target in a gel phantom was 0.35 mm. An accurate punction of the prostate gland can easily be performed using this system with a transgluteal access. The T2-weighted images are superior for the evaluation of the prostate anatomy and the needle position during the interventions. In conclusion, our preliminary results indicate that this MR-guided assistance system is suitable for an accurate transgluteal needle placement in the prostate.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Urol. 1994 Dec;152(6 Pt 2):2304-7 - PubMed
    1. Eur Radiol. 2004 Feb;14(2):318-25 - PubMed
    1. Cancer. 1998 Oct 15;83(8):1679-84 - PubMed
    1. J Urol. 1989 Jul;142(1):76-82 - PubMed
    1. AJR Am J Roentgenol. 1996 Jun;166(6):1332-4 - PubMed

Publication types

LinkOut - more resources