Comparative Study
. 2007 Mar;91(3):313-8.
doi: 10.1136/bjo.2006.096586. Epub 2006 Oct 11.

Accuracy of GDx VCC, HRT I, and clinical assessment of stereoscopic optic nerve head photographs for diagnosing glaucoma

Affiliations
Comparative Study

Accuracy of GDx VCC, HRT I, and clinical assessment of stereoscopic optic nerve head photographs for diagnosing glaucoma

Nicolaas J Reus et al. Br J Ophthalmol. 2007 Mar.

Abstract

Aims: To determine and compare the accuracy and reproducibility of GDx variable cornea compensation (VCC) scanning laser polarimetry (SLP) with VCC, Heidelberg retina tomograph (HRT) I confocal scanning laser ophthalmoscopy (CSLO), and clinical assessment of stereoscopic optic nerve head (ONH) photographs for diagnosing glaucoma.

Methods: One eye each of 40 healthy subjects, 48 glaucoma patients, and six patients with ocular hypertension were measured with SLP-VCC and CSLO. Simultaneous stereoscopic ONH photographs were also obtained. Sixteen photographs of healthy and glaucomatous eyes were duplicated for assessing intraobserver agreement. Four glaucoma specialists, four general ophthalmologists, four residents in ophthalmology, and four optometrists classified the ONH photographs as normal or glaucomatous. For SLP-VCC, the nerve fiber indicator (NFI) was evaluated. For CSLO, the Moorfields regression analysis (MRA) and the Bathija linear discriminant function (LDF) were used. Sensitivity, specificity, percentage of correctly classified eyes, and intra- and interobserver agreement, expressed as kappa (kappa) were calculated.

Results: SLP-VCC had the highest diagnostic accuracy, with a sensitivity, specificity, and overall correct classification of 91.7%, 95.0% and 93.2%, respectively. CSLO, expressed as Bathija LDF and MRA, had a diagnostic accuracy comparable to glaucoma specialists and general ophthalmologists with an overall accuracy of 89.8%, 86.4%, 86.7% and 85.2%, respectively. Residents classified the fewest eyes correctly. Intraobserver agreement for classifying the ONH photographs ranged between 0.48 (within residents) and 0.78 (within glaucoma specialists). The interobserver agreement ranged between 0.45 (between residents) and 0.74 (between glaucoma specialists). The agreement between observers and CSLO MRA (kappa, 0.68) was statistically significantly higher (p<0.001; paired t-test) than between observers and SLP-VCC NFI (kappa, 0.60) and CSLO Bathija LDF (kappa, 0.62).

Conclusion: Automated analysis of measurements with GDx VCC and HRT had a similar diagnostic accuracy for glaucoma as classification of stereoscopic ONH photographs by glaucoma specialists, thus bringing all eye-care professionals to this desirable level. The intra- and interobserver agreement for ONH analysis was only moderate to good. We think these imaging techniques may assist clinicians in diagnosing glaucoma.

PubMed Disclaimer

Conflict of interest statement

Competing interests: NJR was reimbursed for travelling expenses by Laser Diagnostic Technologies, Inc. (now: Carl Zeiss Meditec, Inc.) for speaking at several symposia. MdG has no competing interests. HGL was reimbursed for travelling expenses by Laser Diagnostic Technologies, Inc. (now: Carl Zeiss Meditec, Inc.) for speaking at several symposia and had a research fellow paid for, in part, by Laser Diagnostic Technologies, Inc. (now: Carl Zeiss Meditec, Inc.). NJR and HGL used a GDx VCC that was on loan by Laser Diagnostic Technologies, Inc. (now: Carl Zeiss Meditec, Inc.).

Comment in

References

    1. Varma R, Steinmann W C, Scott I U. Expert agreement in evaluating the optic disc for glaucoma. Ophthalmology 199299215–221. - PubMed
    1. Abrams L S, Scott I U, Spaeth G L.et al Agreement among optometrists, ophthalmologists, and residents in evaluating the optic disc for glaucoma. Ophthalmology 19941011662–1667. - PubMed
    1. Greaney M J, Hoffman D C, Garway‐Heath D F.et al Comparison of optic nerve imaging methods to distinguish normal eyes from those with glaucoma. Invest Ophthalmol Vis Sci 200243140–145. - PubMed
    1. Wollstein G, Garway‐Heath D F, Fontana L.et al Identifying early glaucomatous changes. Comparison between expert clinical assessment of optic disc photographs and confocal scanning ophthalmoscopy. Ophthalmology 20001072272–2277. - PubMed
    1. Girkin C A, McGwin G, Jr, Long C.et al Subjective and objective optic nerve assessment in African Americans and whites. Invest Ophthalmol Vis Sci 2004452272–2278. - PubMed

Publication types