Cardiac perforation 6 weeks after percutaneous atrial septal defect repair using an Amplatzer septal occluder
- PMID: 17043132
- DOI: 10.1542/peds.2005-0825
Cardiac perforation 6 weeks after percutaneous atrial septal defect repair using an Amplatzer septal occluder
Abstract
A 14-year-old boy presented to the emergency department unaccompanied by his parents with a decreased level of consciousness, bradycardia, and hypotension after a syncopal episode. The patient's electronic chart revealed a percutaneous closure of a secundum atrial septal defect using an Amplatzer septal occluder (AGA Medical, Golden Valley, MN) 6 weeks before this presentation. An urgent echocardiogram revealed a moderate pericardial effusion, and 320 mL of fresh blood was evacuated by subxiphoid pericardiocentesis. The child underwent surgical exploration and was found to have a perforation in the superior-posterior aspect of the right atrium, which was corrected. The septal occluder was extracted, and the atrial septal defect was closed with a pericardial patch. This case illustrates a rare but life-threatening complication of percutaneous closure of atrial septal defect using an Amplatzer septal occluder and the importance of timely access to patient records when available history and physical examination are limited.
Similar articles
-
Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder.J Am Coll Cardiol. 2005 Apr 19;45(8):1213-8. doi: 10.1016/j.jacc.2004.12.072. J Am Coll Cardiol. 2005. PMID: 15837251 Review.
-
Early cardiac perforation after atrial septal defect closure with the Amplatzer septal occluder.Ann Thorac Surg. 2006 Jun;81(6):2312-3. doi: 10.1016/j.athoracsur.2005.08.008. Ann Thorac Surg. 2006. PMID: 16731185
-
Aortic arch embolization of an Amplatzer occluder after an atrial septal defect closure: hybrid operative approach without circulatory arrest.Eur J Cardiothorac Surg. 2005 Aug;28(2):340-2. doi: 10.1016/j.ejcts.2005.04.022. Eur J Cardiothorac Surg. 2005. PMID: 15939606
-
Amplatzer septal occluder associated with late pulmonary venous obstruction requiring surgical removal with acquired aorta to left atrial fistula.Am J Cardiol. 2009 Apr 1;103(7):1039-40. doi: 10.1016/j.amjcard.2008.11.057. Epub 2009 Feb 7. Am J Cardiol. 2009. PMID: 19327438
-
Analysis of factors related to successful transcatheter closure of secundum atrial septal defects using the HELEX septal occluder.Am Heart J. 2006 May;151(5):1129.e7-11. doi: 10.1016/j.ahj.2006.01.005. Am Heart J. 2006. PMID: 16644351 Review.
Cited by
-
Analysis of the US Food and Drug Administration Manufacturer and User Facility Device Experience database for adverse events involving Amplatzer septal occluder devices and comparison with the Society of Thoracic Surgery congenital cardiac surgery database.J Thorac Cardiovasc Surg. 2009 Jun;137(6):1334-41. doi: 10.1016/j.jtcvs.2009.02.032. J Thorac Cardiovasc Surg. 2009. PMID: 19464444 Free PMC article.
-
A Rare Ultra-Long-Term Complication of Occluder Recanalization Due to Spontaneous Perforation of Polyvinyl Alcohol Membrane of Atrial Septal Defect Occluder: A Case Report and Review of the Literature.Front Cardiovasc Med. 2022 Jul 22;9:926527. doi: 10.3389/fcvm.2022.926527. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 35935660 Free PMC article.
-
Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.J Artif Organs. 2012 Dec;15(4):311-24. doi: 10.1007/s10047-012-0651-7. Epub 2012 Jun 23. J Artif Organs. 2012. PMID: 22729293 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous