. 2006 Dec;183(2):290-302.
doi: 10.1016/j.jmr.2006.09.004. Epub 2006 Oct 11.

On-line coupling of high temperature GPC and 1H NMR for the analysis of polymers

Affiliations

On-line coupling of high temperature GPC and 1H NMR for the analysis of polymers

Wolf Hiller et al. J Magn Reson. 2006 Dec.

Abstract

The on-line coupling of gel permeation chromatography (GPC) and 1H NMR operating at temperatures up to 130 degrees C is presented. A NMR flow probe with a cell volume of 120 microL and a stop-flow valve are developed for on-flow and stop-flow NMR measurements at high temperatures. To maintain high and constant temperatures through the whole probe, the flow probe contains two separate heating circuits. A modified stop-flow valve is developed as a control device for enabling on-flow and stop-flow experiments at high temperature conditions. Heated transfer lines connect the flow probe with the high temperature GPC system. Due to their semicrystalline nature, polyolefins can be studied by liquid chromatography only at temperatures above 100 degrees C. The novel high temperature GPC-NMR system is used for the separation of complex polyolefins regarding their molar mass and for the analysis of different chemical structures. Blends of polyethylene, poly(methyl methacrylate), and ethylene-methyl methacrylate copolymers are separated according to the molar masses of the components. The compositions of the components are directly studied by on-line NMR. Moreover, the chemical composition distribution of an ethylene-methyl methacrylate copolymer sample is analysed. Differences between results of on-flow and stop-flow measurements are discussed.

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources