A pilot assessment of the FloTrac cardiac output monitoring system
Abstract
Objective: To compare measurement of cardiac output (CO) by means of the FloTrac CO monitor with the pulmonary artery catheter (PAC).
Design: Prospective observational study.
Setting: Intensive care unit of a tertiary hospital.
Patients: Six post-operative cardiac surgery patients with existing arterial cannulas and PACs.
Interventions: Attachment of the FloTrac CO monitor and transducer to an existing arterial cannula. Simultaneous measurements of CO, indexed to body surface area (cardiac index, CI) by the FloTrac CO monitor and by either a bolus thermodilution or continuous CO PAC. Statistical analysis of observations.
Measurements and results: We performed CO measurements in six patients every 1-4 h after cardiac surgery. Comparison of all measurements showed a limited correlation for CI with the two devices (r (2)=0.1218, bias=0.21, 95% limits of agreement -0.81, 1.23). CI measurements obtained with the intermittent bolus PAC had better correlation with the FloTrac CI values (r (2) = 0.2693, bias=-0.0057, 95% limits of agreement -1.2042, 1.1929) than did those obtained with the continuous CO PAC (r (2)=0.0557, bias=0.2436, 95% limits of agreement -0.7350, 1.2222). When analysed according to heart rhythm, CI values measured during atrial pacing showed the best correlation (r (2)=0.377, bias=-0.0244, 95% limits of agreement -0.5226, 0.5714).
Conclusions: CO measurements obtained using the FloTrac CO monitor show a limited correlation with those acquired using the PAC, relatively wide limits of agreement but no clear bias. Further evaluation is required before this device can be recommended for use in the clinical setting.
Similar articles
-
Comparison of FloTrac cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements.Eur J Anaesthesiol. 2007 Oct;24(10):832-9. doi: 10.1017/S0265021507001056. Epub 2007 Aug 1. Eur J Anaesthesiol. 2007. PMID: 17666154 Clinical Trial.
-
Reliability of cardiac output measurements using LiDCOrapid™ and FloTrac/Vigileo™ across broad ranges of cardiac output values.J Clin Monit Comput. 2017 Aug;31(4):709-716. doi: 10.1007/s10877-016-9896-7. Epub 2016 Jun 14. J Clin Monit Comput. 2017. PMID: 27300325 Free PMC article.
-
Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters.Crit Care. 2010;14(6):R212. doi: 10.1186/cc9335. Epub 2010 Nov 23. Crit Care. 2010. PMID: 21092290 Free PMC article.
-
Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac/Vigileo) in haemodynamically unstable patients.Br J Anaesth. 2008 Apr;100(4):451-6. doi: 10.1093/bja/aem409. Epub 2008 Feb 6. Br J Anaesth. 2008. PMID: 18256056 Review.
-
Use of the ClearSight® System for Continuous Noninvasive Hemodynamic Monitoring during Heart Valve Interventions: Review of the Literature and Single-Site Experience.Heart Surg Forum. 2018 Nov 14;21(6):E476-E483. doi: 10.1532/hsf.2177. Heart Surg Forum. 2018. PMID: 30604671 Review.
Cited by
-
Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.Crit Care. 2010;14(3):R109. doi: 10.1186/cc9058. Epub 2010 Jun 10. Crit Care. 2010. PMID: 20537159 Free PMC article.
-
Year in review 2006: Critical Care--Cardiology.Crit Care. 2007;11(4):225. doi: 10.1186/cc5978. Crit Care. 2007. PMID: 17764587 Free PMC article. Review.
-
Continuous arterial pressure waveform monitoring in pediatric cardiac transplant, cardiomyopathy and pulmonary hypertension patients.Intensive Care Med. 2011 Aug;37(8):1297-301. doi: 10.1007/s00134-011-2252-y. Epub 2011 May 31. Intensive Care Med. 2011. PMID: 21626432
-
Intraoperative cardiac function assessment by transesophageal echocardiography versus FloTrac/Vigileo™ system during pectus excavatum surgical repair.J Anesth Analg Crit Care. 2021 Dec 9;1(1):21. doi: 10.1186/s44158-021-00025-4. J Anesth Analg Crit Care. 2021. PMID: 37386660 Free PMC article.
-
Cardiac output monitoring in septic shock: evaluation of the third-generation Flotrac-Vigileo.J Clin Monit Comput. 2013 Jun;27(3):273-9. doi: 10.1007/s10877-013-9431-z. Epub 2013 Jan 30. J Clin Monit Comput. 2013. PMID: 23361128
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources