Review
. 2006 Nov 7;48(9):1896-910.
doi: 10.1016/j.jacc.2006.08.028. Epub 2006 Sep 26.

Diagnostic performance of multislice spiral computed tomography of coronary arteries as compared with conventional invasive coronary angiography: a meta-analysis

Affiliations
Free article
Review

Diagnostic performance of multislice spiral computed tomography of coronary arteries as compared with conventional invasive coronary angiography: a meta-analysis

Michèle Hamon et al. J Am Coll Cardiol. .
Free article

Abstract

Objectives: This study was designed to define the current role of multislice spiral computed tomography (MSCT) for the diagnosis of coronary artery disease (CAD) using a meta-analytic process.

Background: Multislice spiral computed tomography has recently been proposed as an alternative to conventional coronary angiography (CA) for the diagnosis of CAD.

Methods: Using Medline, we identified 29 studies (2,024 patients) evaluating CAD by means of both MSCT (> or =16 slices) and conventional CA before July 2006. After data extraction the analysis was performed according to a random-effects model.

Results: The per-segment analysis pooled the results from 27 studies corresponding to a cumulative number of 22,798 segments. Among unassessable segments, 4.2% were excluded from the analysis and 6.4% were classified at the discretion of the investigators, underscoring the shortcomings of MSCT. With this major limitation, the per-segment sensitivity and specificity were 81% (95% confidence interval [CI] 72% to 89%) and 93% (95% CI 90% to 97%), respectively, with positive and negative likelihood ratios of 21.5 (95% CI 13.1 to 35.5) and 0.11 (95% CI 0.06 to 0.21), respectively, and positive and negative predictive values of 67.8% (95% CI 57.6% to 78.0%) and 96.5% (95% CI 94.7% to 98.3%), respectively. As expected, the per-patient analysis has shown an increased sensitivity of 96% (95% CI 94% to 98%) but a decreased specificity of 74% (95% CI 65% to 84%).

Conclusions: Multislice spiral computed tomography has shortcomings difficult to overcome in daily practice and, at the more clinically relevant per-patient analysis, continues to have moderate specificity in patients with high prevalence of CAD. Studies evaluating the diagnostic performance of the newest generation of MSCT, including patients with low to moderate CAD prevalence, will be critical in establishing the clinical role of this emerging technology as an alternative to CA.