Accurate measurements of hemoglobin oxygen saturation, and fractions of carboxyhemoglobin and methemoglobin in fetal blood using Radiometer OSM3: corrections for fetal hemoglobin fraction and pH
- PMID: 1708524
- DOI: 10.3109/00365519009087516
Accurate measurements of hemoglobin oxygen saturation, and fractions of carboxyhemoglobin and methemoglobin in fetal blood using Radiometer OSM3: corrections for fetal hemoglobin fraction and pH
Abstract
The differences in the visible absorption spectra between fetal and adult oxyhemoglobin and carboxyhemoglobin result in errors in the measurements og hemoglobin oxygen saturation (SO2) and carboxyhemoglobin fraction (FCOHb) in fetal blood, if not corrected for the actual fetal hemoglobin fraction (FHbF) in the sample. In 11 fully oxygenated umbilical cord blood samples (mean FHbF = 77%), we found a mean positive bias in SO2 of 4.7%, and in FCOHb of 2.7%, when measured with a dedicated spectrophotometer (OSM3, Radiometer A/S, Denmark), and using the matrix of absorption coefficients for adult hemoglobin. Accurate measurements were obtained by using OSM3's correction for FHbF in the blood specimen after measurement of FHbF by OSM3. The effects of plasma pH on the measurements of SO2 and FCOHb in fully oxygenated fetal blood were found to be similar to those found for adult blood. From plasma pH 7.05 to 8.02, measured SO2 increased 1.3% and FCOHb 0.6%. Correction for the pH of fetal blood samples should be considered when calibrating OSM3 and in connection with research studies. The effects of FHbF and pH on the measurement of methemoglobin fraction (FMetHb) were less than 0.2%, and can be ignored. FHbF measured by OSM3 at pH 7.4 is about 14% too high compared to alkali denaturation rate method. However, the presence of a metabolic acidemia, which is common in fetal blood specimens, decreases this bias, so that for example in our study, FHbF, measured by OSM3 and uncorrected for pH changes was on average only 6% too high. We recommend that OSM3's factor of 18.6 is reduced to 16.4, and that correction is made for pH.
Similar articles
-
Performance of an automated six-wavelength photometer (Radiometer OSM3) for routine measurement of hemoglobin derivatives.Clin Chem. 1988 Jan;34(1):149-52. Clin Chem. 1988. PMID: 2448064
-
Measured fraction of carboxyhaemoglobin depends on oxygen saturation of haemoglobin.Scand J Clin Lab Invest. 2001 Feb;61(1):83-7. doi: 10.1080/00365510151068063. Scand J Clin Lab Invest. 2001. PMID: 11300615
-
A new trend in blood gas chemistry: the measurement of clinically relevant hemoglobin derivatives. Performance of the OSM3 hemoximeter.Scand J Clin Lab Invest Suppl. 1987;188:57-60. Scand J Clin Lab Invest Suppl. 1987. PMID: 3482481
-
From oxygen content to pulse oximetry: completing the picture in the newborn.Acta Anaesthesiol Scand Suppl. 1995;107:95-100. doi: 10.1111/j.1399-6576.1995.tb04341.x. Acta Anaesthesiol Scand Suppl. 1995. PMID: 8599308 Review.
-
The measurement of dyshemoglobins and total hemoglobin by pulse oximetry.Curr Opin Anaesthesiol. 2008 Dec;21(6):805-10. doi: 10.1097/ACO.0b013e328316bb6f. Curr Opin Anaesthesiol. 2008. PMID: 18997533 Review.
Cited by
-
Methemoglobin determination by multi-component analysis in coho salmon (Oncorhynchus kisutch) possessing unstable hemoglobin.MethodsX. 2020 Feb 22;7:100836. doi: 10.1016/j.mex.2020.100836. eCollection 2020. MethodsX. 2020. PMID: 32257839 Free PMC article.
-
The measurement of accurate fetal hemoglobin and related oxygen saturation by the hemoximeter.Clin Chim Acta. 2006 Dec;374(1-2):75-80. doi: 10.1016/j.cca.2006.05.042. Epub 2006 Jun 6. Clin Chim Acta. 2006. PMID: 16844104 Free PMC article.
-
Validation of oxygen saturation monitoring in neonates.Am J Crit Care. 2007 Mar;16(2):168-78. Am J Crit Care. 2007. PMID: 17322018 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Miscellaneous