Comparison of different techniques of anterior chamber depth and keratometric measurements
- PMID: 17101110
- DOI: 10.1016/j.ajo.2006.08.031
Comparison of different techniques of anterior chamber depth and keratometric measurements
Abstract
Purpose: To compare measurements of anterior chamber depth (ACD) and corneal curvature obtained with the Pentacam to other commonly used devices.
Design: Prospective, nonrandomized clinical trial.
Methods: Measurements of ACD and keratometry were prospectively obtained in 22 eyes of 11 subjects with the Pentacam compared with measurements of the ultrasound (US) A-scan and IOLMaster (ACD), and with those of automated keratometry (AK) and IOLMaster (keratometry) at the Assaf Harofeh Medical Center.
Results: The mean interdevice differences in keratometry for Pentacam vs AK, Pentacam vs IOLMaster, and AK vs IOLMaster were -0.046 diopters, -0.471 diopters, and -0.424 diopters, respectively. Measurements of keratometry by the IOLMaster differed statistically significantly from those of Pentacam (P < .01) and AK (P < .01). For measurement of keratometry, 95% limits of agreement were -1.321 to 1.229 diopters for the Pentacam and AK, -1.478 to 0.536 diopters for the Pentacam and IOLMaster, and -1.026 to 0.178 diopters for the AK and IOLMaster. The mean interdevice differences in ACD for the Pentacam vs US, Pentacam vs IOLMaster, and US vs IOLMaster were 0.103, 0.099, and -0.004 mm, respectively. Measurements of ACD by the Pentacam differed statistically significantly from those of US (P < .05) and IOLMaster (P < .01). For measurement of ACD, 95% limits of agreement were -0.197 to 0.404 mm for the Pentacam and US, -0.169 to 0.367 mm for the Pentacam and IOLMaster, and -0.332 to 0.324 mm for the US and IOLMaster.
Conclusions: For some clinical applications, measurements of ACD and corneal curvature with the tested machines may differ greatly and therefore may not be interchangeable.
Similar articles
-
Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.PLoS One. 2015 Nov 17;10(11):e0143110. doi: 10.1371/journal.pone.0143110. eCollection 2015. PLoS One. 2015. PMID: 26575265 Free PMC article.
-
Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements.Cont Lens Anterior Eye. 2017 Oct;40(5):318-322. doi: 10.1016/j.clae.2017.03.007. Epub 2017 Mar 23. Cont Lens Anterior Eye. 2017. PMID: 28342729
-
A comparison of Scheimpflug imaging simulated and Holladay equivalent keratometry values with partial coherence interferometry keratometry measurements in phakic eyes.Can J Ophthalmol. 2009 Dec;44(6):700-4. doi: 10.3129/i09-172. Can J Ophthalmol. 2009. PMID: 20029491
-
Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.Int J Ophthalmol. 2016 Jul 18;9(7):1057-65. doi: 10.18240/ijo.2016.07.20. eCollection 2016. Int J Ophthalmol. 2016. PMID: 27500117 Free PMC article. Review.
-
Clinical assessment of anterior chamber depth.Ophthalmic Physiol Opt. 1998 Sep;18 Suppl 2:S32-9. doi: 10.1016/s0275-5408(98)00031-3. Ophthalmic Physiol Opt. 1998. PMID: 10793862 Review. No abstract available.
Cited by
-
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices.Clin Ophthalmol. 2017 Jul 12;11:1291-1299. doi: 10.2147/OPTH.S137878. eCollection 2017. Clin Ophthalmol. 2017. PMID: 28744097 Free PMC article.
-
Comparison of axial length, anterior chamber depth and intraocular lens power between IOLMaster and ultrasound in normal, long and short eyes.PLoS One. 2018 Mar 15;13(3):e0194273. doi: 10.1371/journal.pone.0194273. eCollection 2018. PLoS One. 2018. PMID: 29543854 Free PMC article.
-
Update on Intraocular Lens Formulas and Calculations.Asia Pac J Ophthalmol (Phila). 2020 May-Jun;9(3):186-193. doi: 10.1097/APO.0000000000000293. Asia Pac J Ophthalmol (Phila). 2020. PMID: 32501896 Free PMC article. Review.
-
Anterior chamber parameters in cataract surgery candidates from middle China.Medicine (Baltimore). 2022 Dec 9;101(49):e32211. doi: 10.1097/MD.0000000000032211. Medicine (Baltimore). 2022. PMID: 36626461 Free PMC article.
-
Keratometry with five different techniques: a study of device repeatability and inter-device agreement.Int Ophthalmol. 2014 Aug;34(4):869-75. doi: 10.1007/s10792-013-9895-3. Epub 2014 Feb 23. Int Ophthalmol. 2014. PMID: 24562593