Spectral optical coherence tomography: a novel technique for cornea imaging
- PMID: 17102675
- DOI: 10.1097/01.ico.0000224644.81719.59
Spectral optical coherence tomography: a novel technique for cornea imaging
Erratum in
- Cornea. 2007 Jun;26(5):646. Kaluzy, Bartłomiej J [corrected to Kaluzny, Bartłomiej J]
Abstract
Purpose: Spectral optical coherence tomography (SOCT) is a new, noninvasive, noncontact, high-resolution technique that provides cross-sectional images of the objects that weakly absorb and scatter light. SOCT, because of very short acquisition time and high sensitivity, is capable of providing tomograms of substantially better quality than the conventional OCT. The aim of this paper is to show the application of the SOCT to cross-sectional imaging of the cornea and its pathologies.
Methods: Eleven eyes with different corneal pathologies were examined with a slit lamp and the use of a prototype SOCT instrument constructed in the Institute of Physics, Nicolaus Copernicus University, Toruń, Poland.
Results: Our SOCT system provides high-resolution (4 microm axial, 10 microm transversal) tomograms composed of 3000-5000 A-scans with an acquisition time of 120-200 ms. The quality of the images is adequate for detailed cross-sectional evaluation of various corneal pathologies. Objective assessment of the localization, size, shape, and light-scattering properties of the changed tissue is possible. Corneal and epithelial thickness and the depth and width of lesions can be estimated.
Conclusion: SOCT technique allows acquiring clinically valuable cross-sectional optical biopsy of the cornea and its pathologies.
Similar articles
-
Spectral optical coherence tomography: a new imaging technique in contact lens practice.Ophthalmic Physiol Opt. 2006 Mar;26(2):127-32. doi: 10.1111/j.1475-1313.2006.00371.x. Ophthalmic Physiol Opt. 2006. PMID: 16460312
-
Granular corneal dystrophy in 830-nm spectral optical coherence tomography.Cornea. 2008 Aug;27(7):830-2. doi: 10.1097/ICO.0b013e31816be90f. Cornea. 2008. PMID: 18650671
-
Spectral optical coherence tomography in video-rate and 3D imaging of contact lens wear.Optom Vis Sci. 2007 Dec;84(12):1104-9. doi: 10.1097/OPX.0b013e31815b9e0e. Optom Vis Sci. 2007. PMID: 18091301
-
[Application of spectral optical coherent tomography (SOCT) in ophthalmology].Klin Oczna. 2004;106(4-5):699-702. Klin Oczna. 2004. PMID: 15646498 Review. Polish.
-
Anterior segment optical coherence tomography.Semin Ophthalmol. 2010 Sep-Nov;25(5-6):317-23. doi: 10.3109/08820538.2010.518473. Semin Ophthalmol. 2010. PMID: 21091018 Review.
Cited by
-
Design of a Linear Wavenumber Spectrometer for Line Scanning Optical Coherence Tomography with 50 mm Focal Length Cylindrical Optics.Sensors (Basel). 2022 Apr 25;22(9):3278. doi: 10.3390/s22093278. Sensors (Basel). 2022. PMID: 35590968 Free PMC article.
-
Use of ultra-high-resolution optical coherence tomography to detect in vivo characteristics of Descemet's membrane in Fuchs' dystrophy.Ophthalmology. 2010 Jun;117(6):1220-7. doi: 10.1016/j.ophtha.2009.10.027. Epub 2010 Feb 16. Ophthalmology. 2010. PMID: 20163865 Free PMC article.
-
Anterior lens capsule and epithelium thickness measurements using spectral-domain optical coherence tomography.BMC Ophthalmol. 2017 Jun 19;17(1):94. doi: 10.1186/s12886-017-0489-0. BMC Ophthalmol. 2017. PMID: 28629391 Free PMC article.
-
Accommodative changes in human eye observed by Kitasato anterior segment optical coherence tomography.Jpn J Ophthalmol. 2013 Jan;57(1):113-9. doi: 10.1007/s10384-012-0208-6. Epub 2012 Nov 21. Jpn J Ophthalmol. 2013. PMID: 23179763
-
Reliability of RTVue, Visante, and slit-lamp adapted ultrasonic pachymetry for central corneal thickness measurement.Yonsei Med J. 2012 May;53(3):634-41. doi: 10.3349/ymj.2012.53.3.634. Yonsei Med J. 2012. PMID: 22477010 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical