Assessment of a race-specific normative HRT-III database to differentiate glaucomatous from normal eyes
- PMID: 17106370
- DOI: 10.1097/01.ijg.0000212289.00917.a8
Assessment of a race-specific normative HRT-III database to differentiate glaucomatous from normal eyes
Abstract
Purpose: To determine if a new, normative, race-specific database enhances the ability of confocal scanning laser ophthalmoscopy to differentiate normal from glaucomatous eyes.
Methods: One eye of eligible normal and glaucoma patients was enrolled. All subjects underwent a complete ophthalmologic examination, standard achromatic perimetry (SITA-SAP, 24-2), and confocal scanning laser ophthalmoscopy [Heidelberg retinal tomograph (HRT-II)] within 1 month of enrollment. Racial groups were defined by self-report. Glaucoma was defined by the existence of reproducible SAP loss (pattern standard deviation <5% and/or Glaucoma Hemifield Test outside normal limits) on 2 consecutive fields. Normal subjects had 2 normal visual fields (pattern standard deviation >5% and Glaucoma Hemifield Test within 97% normal limits) and a normal clinical examination. HRT-II examinations were exported to the HRT-III software, which includes a large race-specific normative database consisting of 733 white and 215 black eyes. Moorfields regression analysis (MRA) for the most abnormal optic disc sector was compared between the HRT-II (MRA2) and the HRT-III software before (MRA3-B) and after (MRA3-A) adjustment for race. Sectors outside the 99.9% confidence interval limits ("outside normal limits") were determined to be abnormal.
Results: We enrolled 124 black (52 glaucoma, 72 normal) and 96 white (32 glaucoma, 64 normal) subjects. Mean age was 51+/-13 years and 50+/-16 years for blacks and whites, respectively (P = 0.45). Visual field mean deviation was -7.3+/-6.7 db for glaucomatous eyes and -0.4+/-1.1 db for normal eyes (P < 0.001). Sensitivity and specificity for the HRT-II was 71.9% and 95.3%, respectively, for white subjects and 50.0% and 98.6%, respectively, for black subjects. Using the expanded HRT-III database, analysis yielded a sensitivity of 81.3% and specificity of 93.8% for whites and a sensitivity of 71.2% and specificity of 86.1% for blacks. After an adjustment for black ethnicity was made in the HRT-III program, the sensitivity and specificity for blacks was 65.4% and 90.3%, respectively.
Conclusions: A new, larger, race-specific HRT-III database increases sensitivity while maintaining specificity for whites and increases sensitivity but decreases specificity for blacks. New software and databases based on race require careful scrutiny before use in clinical practice.
Similar articles
-
Detection of glaucoma using operator-dependent versus operator-independent classification in the Heidelberg retinal tomograph-III.Br J Ophthalmol. 2006 Nov;90(11):1390-2. doi: 10.1136/bjo.2006.098111. Epub 2006 Jul 26. Br J Ophthalmol. 2006. PMID: 16870651 Free PMC article.
-
Comparison of diagnostic accuracy of Heidelberg Retina Tomograph II and Heidelberg Retina Tomograph 3 to discriminate glaucomatous and nonglaucomatous eyes.Am J Ophthalmol. 2007 Oct;144(4):525-32. doi: 10.1016/j.ajo.2007.06.021. Epub 2007 Aug 13. Am J Ophthalmol. 2007. PMID: 17693382 Free PMC article.
-
Detection of glaucomatous visual field changes using the Moorfields regression analysis of the Heidelberg retina tomograph.Am J Ophthalmol. 2003 Jul;136(1):26-33. doi: 10.1016/s0002-9394(03)00084-9. Am J Ophthalmol. 2003. PMID: 12834666
-
New developments in Heidelberg retina tomograph for glaucoma.Curr Opin Ophthalmol. 2008 Mar;19(2):141-8. doi: 10.1097/ICU.0b013e3282f4450b. Curr Opin Ophthalmol. 2008. PMID: 18301288 Review.
-
Glaucoma and race: a case for screening in Connecticut.Conn Med. 2000 Feb;64(2):75-8. Conn Med. 2000. PMID: 10743748 Review.
Cited by
-
Variation of laminar depth in normal eyes with age and race.Invest Ophthalmol Vis Sci. 2014 Nov 20;55(12):8123-33. doi: 10.1167/iovs.14-15251. Invest Ophthalmol Vis Sci. 2014. PMID: 25414182 Free PMC article.
-
Optic nerve head and fibre layer imaging for diagnosing glaucoma.Cochrane Database Syst Rev. 2015 Nov 30;2015(11):CD008803. doi: 10.1002/14651858.CD008803.pub2. Cochrane Database Syst Rev. 2015. PMID: 26618332 Free PMC article.
-
Effect of disease severity and optic disc size on diagnostic accuracy of RTVue spectral domain optical coherence tomograph in glaucoma.Invest Ophthalmol Vis Sci. 2011 Mar 10;52(3):1290-6. doi: 10.1167/iovs.10-5546. Print 2011 Mar. Invest Ophthalmol Vis Sci. 2011. PMID: 20811060 Free PMC article.
-
Bruch's Membrane Opening Minimum Rim Width and Retinal Nerve Fiber Layer Thickness in a Normal White Population: A Multicenter Study.Ophthalmology. 2015 Sep;122(9):1786-94. doi: 10.1016/j.ophtha.2015.06.001. Epub 2015 Jul 18. Ophthalmology. 2015. PMID: 26198806 Free PMC article.
-
HRT for the Diagnosis and Detection of Glaucoma Progression.Open Ophthalmol J. 2015 May 15;9:58-67. doi: 10.2174/1874364101509010058. eCollection 2015. Open Ophthalmol J. 2015. PMID: 26069518 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous