Multislice computed tomography angiography screening for new aneurysms in patients with previously clip-treated intracranial aneurysms: Feasibility, positive predictive value, and interobserver agreement
- PMID: 17121128
- DOI: 10.3171/jns.2006.105.5.682
Multislice computed tomography angiography screening for new aneurysms in patients with previously clip-treated intracranial aneurysms: Feasibility, positive predictive value, and interobserver agreement
Abstract
Object: Multislice computed tomography (CT) angiography may be useful for screening patients with intracranial aneurysms that are treated with clip occlusion. However, cobalt clips produce much more artifact on CT scans than titanium clips, which may hamper the evaluation of the image obtained at the clip site.
Methods: The authors screened 415 patients with previously ruptured aneurysms that had been treated using cobalt clips. Screening was performed using multislice CT angiography. The feasibility of this modality for screening these patients (based on the complication risk, CT angiography quality, and artifact avoidance) and interobserver agreement were evaluated. Patients in whom the presence of an aneurysm was suspected based on results of CT angiography studies underwent digital subtraction (DS) angiography. False-negative and false-positive findings were recorded, and the positive predictive value (PPV) was calculated. Eight patients (1.9%) had allergies to the contrast material. The quality of the CT angiography image was suboptimal in 14%. In 52%, clip artifacts hampered evaluation of the clip site. In 65 patients who underwent DS angiography, there were nine false-positive and eight false-negative reports related to aneurysms that were either small, located at the clip site, or were infundibula. The PPV on a per-patient basis was 86% (95% confidence interval [CI] 75-94%); for aneurysms at the clip site it was 83% (95% CI 61-95%); and for aneurysms at different locations it was 91% (95% CI 81-97%). The interobserver agreement was good (kappa = 0.69; 95% CI 0.60-0.78).
Conclusions: Except for the evaluation of images from the clip site, CT angiography has good feasibility with good PPV and interobserver agreement. Drawbacks are that very small aneurysms can be missed and that visualization is poor at the clip site in patients in whom cobalt clips have been placed for occlusion. This second problem can be expected to resolve with the increasing use of titanium clips.
Similar articles
-
Comparison of multislice computerized tomography angiography and digital subtraction angiography in the postoperative evaluation of patients with clipped aneurysms.J Neurosurg. 2006 Mar;104(3):395-403. doi: 10.3171/jns.2006.104.3.395. J Neurosurg. 2006. PMID: 16572652 Clinical Trial.
-
Limitations of three-dimensional reconstructed computerized tomography angiography after clip placement for intracranial aneurysms.J Neurosurg. 2005 Oct;103(4):656-61. doi: 10.3171/jns.2005.103.4.0656. J Neurosurg. 2005. PMID: 16266048
-
Sixteen-row multislice computed tomography angiography in the diagnosis and characterization of intracranial aneurysms: comparison with conventional angiography and intraoperative findings.J Neurosurg. 2008 Jun;108(6):1184-91. doi: 10.3171/JNS/2008/108/6/1184. J Neurosurg. 2008. PMID: 18518726 Clinical Trial.
-
Image quality and artefact generation post-cerebral aneurysm clipping using a 64-row multislice computer tomography angiography (MSCTA) technology: A retrospective study and review of the literature.Clin Neurol Neurosurg. 2010 Jun;112(5):386-91. doi: 10.1016/j.clineuro.2010.02.001. Epub 2010 Mar 1. Clin Neurol Neurosurg. 2010. PMID: 20189713 Review.
-
Virtues and drawbacks of titanium alloy aneurysm clips.Acta Neurochir Suppl. 1999;72:81-8. doi: 10.1007/978-3-7091-6377-1_7. Acta Neurochir Suppl. 1999. PMID: 10337415 Review.
Cited by
-
Aneurysmal subarachnoid haemorrhage from a neuroimaging perspective.Crit Care. 2014 Nov 13;18(6):557. doi: 10.1186/s13054-014-0557-2. Crit Care. 2014. PMID: 25673429 Free PMC article. Review.
-
Subtraction 3D CT angiography with the orbital synchronized helical scan technique for the evaluation of postoperative cerebral aneurysms treated with cobalt-alloy clips.AJNR Am J Neuroradiol. 2008 Jun;29(6):1071-5. doi: 10.3174/ajnr.A1040. Epub 2008 Mar 27. AJNR Am J Neuroradiol. 2008. PMID: 18372416 Free PMC article.
-
Long-term follow-up survey reveals a high yield, up to 30% of patients presenting newly detected aneurysms more than 10 years after ruptured intracranial aneurysms clipping.Neurosurg Rev. 2011 Oct;34(4):485-96. doi: 10.1007/s10143-011-0332-7. Epub 2011 Jun 4. Neurosurg Rev. 2011. PMID: 21643681
-
The value of dual-energy CTA for control of surgically clipped aneurysms.Eur Radiol. 2011 Oct;21(10):2193-201. doi: 10.1007/s00330-011-2147-x. Epub 2011 May 10. Eur Radiol. 2011. PMID: 21556907
-
Applications of multislice CT angiography in the surgical clipping and endovascular coiling of intracranial aneurysms.J Biomed Res. 2010 Nov;24(6):467-73. doi: 10.1016/S1674-8301(10)60062-0. J Biomed Res. 2010. PMID: 23554664 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical