Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor
- PMID: 17130198
- DOI: 10.2337/dc06-1361
Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor
Abstract
Objective: In this study, we evaluated the safety and efficacy of 7-day transcutaneous, real-time, continuous glucose monitoring (CGM) in subjects with insulin-requiring diabetes.
Research design and methods: Eighty-six subjects were enrolled at five U.S. centers. Subjects wore a sensor inserted under the skin of the abdomen for 7 days during each of three consecutive periods. Data were blinded during period 1 and unblinded during periods 2 and 3.
Results: Of the 6,811 matched self-monitoring of blood glucose to sensor values prospectively analyzed, 97.2% fell in the Clarke error grid zones A and B, and median absolute relative difference was 11.4%. After unblinding, subjects reduced time spent at <55 mg/dl by 0.3 h/day, reduced time spent at >240 mg/dl by 1.5 h/day, and increased time in the target zone (81-140 mg/dl) by 1.4 h/day (P < 0.05 for all three comparisons). Improvements were seen in both types 1 and 2 diabetes and with use of both multiple daily injections and continuous subcutaneous insulin infusion. Modal day graphs were generated in six groups of subjects based on HbA1c (A1C) (<or=6, 6-7, 7-8, 8-9, 9-10, and >10%). Mean glucose levels from midnight to 7:00 a.m. (fasting and dawn phenomenon periods) were only normal for subjects with A1C <or=6%. All other groups were hyperglycemic during this and all periods. Reductions in overall mean glucose were achieved for the four highest A1C groupings with unblinded device use.
Conclusions: This is the first report of a real-time, transcutaneous glucose sensor that functioned for 7 days. The use of CGM in the unblinded phase resulted in improvements in target-range glycemia across all A1C values.
Similar articles
-
Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial.Diabetes Care. 2006 Jan;29(1):44-50. doi: 10.2337/diacare.29.01.06.dc05-1686. Diabetes Care. 2006. PMID: 16373894 Clinical Trial.
-
The effect of congestive heart failure on sensor accuracy among hospitalized patients with type 2 diabetes.Diabetes Technol Ther. 2013 Oct;15(10):817-24. doi: 10.1089/dia.2013.0094. Diabetes Technol Ther. 2013. PMID: 24050738 Free PMC article.
-
Accuracy assessment of online glucose monitoring by a subcutaneous enzymatic glucose sensor during exercise in patients with type 1 diabetes treated by continuous subcutaneous insulin infusion.Diabetes Metab. 2013 May;39(3):258-62. doi: 10.1016/j.diabet.2012.12.004. Epub 2013 Mar 19. Diabetes Metab. 2013. PMID: 23522730
-
Accuracy of Continuous Glucose Monitoring During Three Closed-Loop Home Studies Under Free-Living Conditions.Diabetes Technol Ther. 2015 Nov;17(11):801-7. doi: 10.1089/dia.2015.0062. Epub 2015 Aug 4. Diabetes Technol Ther. 2015. PMID: 26241693 Free PMC article. Clinical Trial.
-
Incremental value of continuous glucose monitoring when starting pump therapy in patients with poorly controlled type 1 diabetes: the RealTrend study.Diabetes Care. 2009 Dec;32(12):2245-50. doi: 10.2337/dc09-0750. Epub 2009 Sep 18. Diabetes Care. 2009. PMID: 19767384 Free PMC article. Clinical Trial.
Cited by
-
A new-generation continuous glucose monitoring system: improved accuracy and reliability compared with a previous-generation system.Diabetes Technol Ther. 2013 Oct;15(10):881-8. doi: 10.1089/dia.2013.0077. Epub 2013 Jun 18. Diabetes Technol Ther. 2013. PMID: 23777402 Free PMC article.
-
Glycated Hemoglobin Differences Among Blog-Reading Adults With Type 1 Diabetes Compared With Those Who Do Not Read Blogs: Cross-Sectional Study.JMIR Diabetes. 2019 Apr 2;4(2):e13634. doi: 10.2196/13634. JMIR Diabetes. 2019. PMID: 30938693 Free PMC article.
-
Feasibility of 10-day use of a continuous glucose-monitoring system in adults with type 1 diabetes.Diabetes Care. 2009 Mar;32(3):436-8. doi: 10.2337/dc08-1745. Epub 2008 Nov 25. Diabetes Care. 2009. PMID: 19033411 Free PMC article.
-
Recommendations for Using Real-Time Continuous Glucose Monitoring (rtCGM) Data for Insulin Adjustments in Type 1 Diabetes.J Diabetes Sci Technol. 2017 Jan;11(1):138-147. doi: 10.1177/1932296816663747. Epub 2016 Aug 20. J Diabetes Sci Technol. 2017. PMID: 27530720 Free PMC article. Review.
-
Telemedical artificial pancreas: PARIS (Pancreas Artificial Telemedico Inteligente) research project.Diabetes Care. 2009 Nov;32 Suppl 2(Suppl 2):S211-6. doi: 10.2337/dc09-S313. Diabetes Care. 2009. PMID: 19875554 Free PMC article. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical