Evaluation of a new soft tipped injector for the implantation of foldable intraocular lenses
- PMID: 17135340
- PMCID: PMC1954791
- DOI: 10.1136/bjo.2006.103192
Evaluation of a new soft tipped injector for the implantation of foldable intraocular lenses
Abstract
Aim: To evaluate the R-INJ-04 soft-tipped injector, a new injector with an integral round nozzle manufactured by Rayner Intraocular Lenses, England.
Methods: 16 Rayner C-flex intraocular lenses (IOLs; Rayner Intraocular lenses, England) ranging between +10 and +30 D (2 for each power) were tested. An ophthalmic viscoelastic device (Healon, AMO, Santa Ana, California, USA) was applied to the injectors. The IOLs were loaded according to the company injector's instructions for use and were injected into a Petri dish. After the injection, all the IOLs and nozzles were evaluated by gross (macroscopic) and microscopic analyses and photographed under a light microscope. One lens of each power and the cartridge used for the implantation were then sent for further analysis by scanning electron microscopy (SEM). The rest of the IOLs were tested for power and modulation transfer function (MTF).
Results: All the injections were successful. No damage to the IOLs or to the injectors was found by gross examination, light microscopy and SEM. No deposits were found on the IOL optical surfaces or haptics. Power and MTF analysis showed a close match with the original measurements.
Conclusion: Our results suggest that the R-INJ-04 soft-tipped injector is safe for the implantation of the C-flex IOL with power range from 10 to 30 D. No structural damage to the IOLs or to the injectors was found, and the lens power and light transmission properties were not damaged in any way by the injection process.
Conflict of interest statement
Competing interests: The authors have no financial or proprietary interest in any product mentioned in this paper. DJA is a consultant to Rayner Intraocular lenses, England.
References
-
- Milazzo S, Turut P, Blin H. Alterations to the AcrySof intraocular lens during folding. J Cataract Refract Surg 1996221351–1354. - PubMed
-
- Vrabec M P, Syverud J C, Burgess C J. Forceps‐induced scratching of a foldable acrylic intraocular lens [letter]. Arch Ophthalmol 1996114777 - PubMed
-
- Pfister D R. Stress fractures after folding an acrylic intraocular lens. Am J Ophthalmol 1996121572–574. - PubMed
-
- Carlson K H, Johnson D W. Cracking of acrylic intraocular lenses during capsular bag insertion. Ophthalmic Surg Lasers 199526572–573. - PubMed
-
- Kohnen T, Magdowski G, Koch D D. Scanning electron microscopic analysis of foldable acrylic and hydrogel intraocular lenses. J Cataract Refract Surg 1996221342–1350. - PubMed