A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation
- PMID: 17136486
- DOI: 10.1007/s00198-006-0280-1
A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation
Abstract
Introduction: Short-term precision is often quoted and used as the most important performance parameter of a dual-energy X-ray absorptiometry (DXA) scanner; however, long-term precision has a more profound impact on patient monitoring. Long-term precision refers to the combination of in-vivo precision errors and long-term equipment stability.
Methods: To monitor long-term equipment stability, a phantom was designed with four inserts ranging in bone-mineral density from 0.5 to 3.3 g/cm(2). This phantom was used to monitor the equipment stability of four modern fan-beam densitometers, two each from Hologic and GE/Lunar, over a 4-year period. Manufacturer-recommended quality assurance (QA) procedures were performed, and the scanners stayed within manufacturer-specified tolerances throughout the study.
Results and conclusion: During the 4-year period, the Hologic scanners were observed to cause clinically insignificant BMD shifts (maximum of 0.34%), whereas the GE/Lunar scanners revealed BMD shifts that were clinically significant (1.5% and 2.1%). As a result, using least-significant-change (LSC) calculations based only on short-term in-vivo precision studies for monitoring patients is not valid for the two GE/Lunar densitometers due to the poorer long-term stability they exhibited.
Similar articles
-
Accuracy and precision of 62 bone densitometers using a European Spine Phantom.Osteoporos Int. 1999;10(1):14-9. doi: 10.1007/s001980050188. Osteoporos Int. 1999. PMID: 10501774
-
A comparison of phantoms for cross-calibration of lumbar spine DXA.Osteoporos Int. 2002 Dec;13(12):948-54. doi: 10.1007/s001980200132. Osteoporos Int. 2002. PMID: 12459937
-
Comparison of BMD precision for Prodigy and Delphi spine and femur scans.Osteoporos Int. 2006;17(9):1303-8. doi: 10.1007/s00198-006-0127-9. Epub 2006 Jul 6. Osteoporos Int. 2006. PMID: 16823544
-
Cross-calibration, Least Significant Change and Quality Assurance in Multiple Dual-Energy X-ray Absorptiometry Scanner Environments: 2019 ISCD Official Position.J Clin Densitom. 2019 Oct-Dec;22(4):472-483. doi: 10.1016/j.jocd.2019.09.001. Epub 2019 Sep 7. J Clin Densitom. 2019. PMID: 31558404 Review.
-
The impact of recent technological advances on the trueness and precision of DXA to assess body composition.Obesity (Silver Spring). 2012 Jan;20(1):30-9. doi: 10.1038/oby.2011.211. Epub 2011 Jul 14. Obesity (Silver Spring). 2012. PMID: 21760631 Review.
Cited by
-
Velaglucerase alfa (VPRIV) enzyme replacement therapy in patients with Gaucher disease: Long-term data from phase III clinical trials.Am J Hematol. 2015 Jul;90(7):584-91. doi: 10.1002/ajh.24012. Am J Hematol. 2015. PMID: 25801797 Free PMC article. Clinical Trial.
-
Geographic variation in bone mineral density and prevalent fractures in the Canadian longitudinal study on aging.Osteoporos Int. 2024 Apr;35(4):599-611. doi: 10.1007/s00198-023-06975-5. Epub 2023 Dec 2. Osteoporos Int. 2024. PMID: 38040857
-
Optimal monitoring time interval between DXA measures in children.J Bone Miner Res. 2011 Nov;26(11):2745-52. doi: 10.1002/jbmr.473. J Bone Miner Res. 2011. PMID: 21773995 Free PMC article.
-
Does a spinal implant alter dual energy X-ray absorptiometry body composition measurements?PLoS One. 2019 Sep 19;14(9):e0222758. doi: 10.1371/journal.pone.0222758. eCollection 2019. PLoS One. 2019. PMID: 31536597 Free PMC article.
-
Cortical bone mechanics technology signal quality maintains robustness across a range of biometric profiles.JBMR Plus. 2025 Jul 9;9(9):ziaf116. doi: 10.1093/jbmrpl/ziaf116. eCollection 2025 Sep. JBMR Plus. 2025. PMID: 40842906 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials