Objective: Injection of intraarticular corticosteroid remains an important therapy for inflammatory arthritis. In a randomized controlled trial we compared the new reciprocating procedure device (RPD) to the traditional syringe for injection of intraarticular corticosteroid.
Methods: One hundred fifty-four intraarticular corticosteroid injection procedures were randomized to the conventional syringe or the RPD. Using the syringe or RPD, the needle was introduced into the joint, any effusion that was present was aspirated, and the corticosteroid (methylprednisolone acetate) was injected. Outcome measures included patient pain measured by visual analog scale (VAS pain), procedure duration, operator satisfaction, complications, and immediate and delayed response to the injected medication.
Results: The RPD reduced pain scores by 49% (RPD VAS pain score: 2.40 +/- 2.17; conventional syringe VAS pain score: 4.73 +/- 3.39; p < 0.001), reduced procedure time by 31% (RPD: 1.28 +/- 1.08 min, conventional syringe: 1.86 +/- 1.26; p < 0.01), and improved physician satisfaction with the joint procedure device by 63% (RPD visual analog satisfaction scale score: 9.12 +/- 0.80, conventional syringe 5.59 +/- 1.28; p < 0.001). Fifty-five percent (43/78) of patients experienced moderate to severe pain (VAS pain > or = 5) with the conventional syringe, while 17% (13/76) experienced moderate to severe pain with the RPD. The same beneficial response was present when intermediate or large joints were analyzed separately. Longterm outcomes were equivalent.
Conclusion: When a conventional syringe is used for corticosteroid injection, many patients experience significant procedural pain. The RPD significantly reduces patient pain, reduces procedure time, and improves operator satisfaction. The RPD is superior to the traditional syringe for injection of intraarticular corticosteroid.
Comment in
-
Intra-articular corticosteroid injections with the reciprocating procedure device reduced procedural pain and duration more than the conventional syringe.Evid Based Med. 2007 Aug;12(4):106. doi: 10.1136/ebm.12.4.106. Evid Based Med. 2007. PMID: 17885154 No abstract available.
Similar articles
-
Reduced pain of intraarticular hyaluronate injection with the reciprocating procedure device.J Clin Rheumatol. 2007 Feb;13(1):16-9. doi: 10.1097/01.rhu.0000256280.85507.bd. J Clin Rheumatol. 2007. PMID: 17278943 Clinical Trial.
-
A randomized, controlled trial of the reciprocating procedure device for local anesthesia.J Emerg Med. 2008 Aug;35(2):119-25. doi: 10.1016/j.jemermed.2007.08.060. Epub 2008 Feb 20. J Emerg Med. 2008. PMID: 18281176 Clinical Trial.
-
Integration of safety technologies into rheumatology and orthopedics practices: a randomized, controlled trial.Arthritis Rheum. 2008 Jul;58(7):1907-14. doi: 10.1002/art.23499. Arthritis Rheum. 2008. PMID: 18576358 Clinical Trial.
-
Prospective, double-blind, randomized placebo-controlled trials in interventional spine: what the highest quality literature tells us.Spine J. 2009 Aug;9(8):690-703. doi: 10.1016/j.spinee.2008.06.447. Epub 2008 Sep 11. Spine J. 2009. PMID: 18789773 Review.
-
Injectable corticosteroids in treatment of arthritis of the knee.Am J Med. 2005 Nov;118(11):1208-14. doi: 10.1016/j.amjmed.2005.05.003. Am J Med. 2005. PMID: 16271901 Review.
Cited by
-
Outcomes and cost-effectiveness of carpal tunnel injections using sonographic needle guidance.Clin Rheumatol. 2014 Jun;33(6):849-58. doi: 10.1007/s10067-013-2438-5. Epub 2013 Nov 26. Clin Rheumatol. 2014. PMID: 24277115 Clinical Trial.
-
Image-directed fine-needle aspiration biopsy of the thyroid with safety-engineered devices.Cardiovasc Intervent Radiol. 2011 Oct;34(5):1006-13. doi: 10.1007/s00270-010-0013-x. Epub 2010 Nov 6. Cardiovasc Intervent Radiol. 2011. PMID: 21057794 Free PMC article.
-
The highly accurate anteriolateral portal for injecting the knee.Sports Med Arthrosc Rehabil Ther Technol. 2011 Mar 30;3(1):6. doi: 10.1186/1758-2555-3-6. Sports Med Arthrosc Rehabil Ther Technol. 2011. PMID: 21447197 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical