Evaluation of the accuracy of a wireless pressure sensor in a canine model of retrograde-collateral (type II) endoleak and correlation with histologic analysis
- PMID: 17145435
- DOI: 10.1016/j.jvs.2006.08.031
Evaluation of the accuracy of a wireless pressure sensor in a canine model of retrograde-collateral (type II) endoleak and correlation with histologic analysis
Abstract
Background: The utility of intra-aneurysmal pressure determination is dependent on the ability to measure pressure in the presence of endoleak and thrombosis. In this study, the accuracy of a CardioMEMS wireless pressure sensor (CardioMEMS, Atlanta, Ga) transducer in the presence of thrombus associated with type II endoleak was measured.
Methods: Type II endoleaks were created in four mongrel dogs by implanting four collateral arterial side branches (lumbar and caudal mesenteric arteries) as a Carrel patch onto a 3-cm prosthetic polytetrafluoroethylene abdominal aortic aneurysm (AAA). The aneurysm was excluded 2 weeks later from antegrade perfusion by a stent graft. The wireless pressure sensor was positioned in the AAA external to the stent graft. A Konigsberg intraluminal solid-state strain-gauge pressure transducer (Konigsberg Instruments, Pasadena, Calif) that is accurate in the presence of thrombus served as the control to determine AAA pressure. Both of the transducers were implanted on the luminal surface of the aneurysm, 180 degrees opposite from the Carrel patch and endoleak channel. Intra-aneurysmal pressure resulting from the type II endoleak was measured twice daily for 4 weeks using both transducers. A total of 56 pre-exclusion and 224 post-exclusion distinct pressure determinations were made. Intra-aneurysmal pressure was indexed to the systemic pressure that was simultaneously measured by a strain-gauge pressure transducer implanted in the native aorta. Histologic analysis of the aneurysm contents was performed with hematoxylin and eosin.
Results: The intra-aneurysmal systolic, mean, and pulse pressures produced by the type II endoleak were significantly lower than systemic pressure in all animals and were < 60% of systemic pressure (P < .001). Close correlation between the wireless transducer and the control strain-gauge transducer was observed (R = 0.83, P < .001). Arteriography and Doppler ultrasound documented retrograde flow through the aneurysm side branches and persistent endoleak patency up to the time of euthanasia. Pathologic analysis demonstrated the endoleak channel to be patent and separated from the transducers by thrombus, which surrounded both transducers.
Conclusions: Intra-aneurysmal pressure generated by type II endoleaks may be accurately measured through thrombus using a wireless pressure sensor in the canine model. The wireless sensor has the potential for clinical applicability in diagnosing and characterizing type II endoleaks.
Similar articles
-
Characterization of retrograde collateral (type II) endoleak using a new canine model.J Vasc Surg. 2004 Nov;40(5):985-94. doi: 10.1016/j.jvs.2004.07.049. J Vasc Surg. 2004. PMID: 15557915
-
Treatment of type II endoleaks with a novel polyurethane thrombogenic foam: induction of endoleak thrombosis and elimination of intra-aneurysmal pressure in the canine model.J Vasc Surg. 2005 Aug;42(2):321-8. doi: 10.1016/j.jvs.2005.04.043. J Vasc Surg. 2005. PMID: 16102634
-
A canine model to study the significance and hemodynamics of type II endoleaks.J Surg Res. 2005 Feb;123(2):275-83. doi: 10.1016/j.jss.2004.08.022. J Surg Res. 2005. PMID: 15680390
-
Evaluation of pressure transmission and intra-aneurysmal contents after endovascular repair using the Trivascular Enovus expanded polytetrafluoroethylene stent graft in a canine model of abdominal aortic aneurysm.J Vasc Surg. 2007 Nov;46(5):1005-13. doi: 10.1016/j.jvs.2007.06.041. Epub 2007 Oct 1. J Vasc Surg. 2007. PMID: 17905556
-
Models of abdominal aortic aneurysm: characterization and clinical applications.Vascular. 2006 Nov-Dec;14(6):343-52. doi: 10.2310/6670.2006.00059. Vascular. 2006. PMID: 17150155 Review.
Cited by
-
Aortic aneurysm sac pressure measurements after endovascular repair using an implantable remote sensor: initial experience and short-term follow-up.Eur Radiol. 2008 May;18(5):957-65. doi: 10.1007/s00330-007-0831-7. Epub 2007 Dec 20. Eur Radiol. 2008. PMID: 18094972
-
Design, fabrication, and implementation of a wireless, passive implantable pressure sensor based on magnetic higher-order harmonic fields.Biosensors (Basel). 2011 Oct 17;1(4):134-52. doi: 10.3390/bios1040134. Biosensors (Basel). 2011. PMID: 25585564 Free PMC article.
-
Implantable Pressure-Sensing Devices for Monitoring Abdominal Aortic Aneurysms in Post-Endovascular Aneurysm Repair.Sensors (Basel). 2024 May 30;24(11):3526. doi: 10.3390/s24113526. Sensors (Basel). 2024. PMID: 38894317 Free PMC article. Review.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical