Comparison of inverse-planned three-dimensional conformal radiotherapy and intensity-modulated radiotherapy for non-small-cell lung cancer
- PMID: 17187941
- DOI: 10.1016/j.ijrobp.2006.09.047
Comparison of inverse-planned three-dimensional conformal radiotherapy and intensity-modulated radiotherapy for non-small-cell lung cancer
Abstract
Purpose: Lungs are the major dose-limiting organ during radiotherapy (RT) for non-small-cell lung cancer owing to the development of pneumonitis. This study compared intensity-modulated RT (IMRT) with three-dimensional conformal RT (3D-CRT) in reducing the dose to the lungs.
Methods: Ten patients with localized non-small-cell lung cancer underwent computed tomography (CT). The planning target volume (PTV) was defined and the organs at risk were outlined. An inverse-planning program, AutoPlan, was used to design the beam angle-optimized six-field noncoplanar 3D-CRT plans. Each 3D-CRT plan was compared with a series of five IMRT plans per patient. The IMRT plans were created using a commercial algorithm and consisted of a series of three, five, seven, and nine equidistant coplanar field arrangements and one six-field noncoplanar plan. The planning objectives were to minimize the lung dose while maintaining the dose to the PTV. The percentage of lung volume receiving >20 Gy (V20) and the percentage of the PTV covered by the 90% isodose (PTV90) were the primary endpoints. The PTV90/V20 ratio was used as the parameter accounting for both the reduction in lung volume treated and the PTV coverage.
Results: All IMRT plans, except for the three-field coplanar plans, improved the PTV90/V20 ratio significantly compared with the optimized 3D-CRT plan. Nine coplanar IMRT beams were significantly better than five or seven coplanar IMRT beams, with an improved PTV90/V20 ratio.
Conclusion: The results of our study have shown that IMRT can reduce the dose to the lungs compared with 3D-CRT by improving the conformity of the plan.
Similar articles
-
A potential to reduce pulmonary toxicity: the use of perfusion SPECT with IMRT for functional lung avoidance in radiotherapy of non-small cell lung cancer.Radiother Oncol. 2007 May;83(2):156-62. doi: 10.1016/j.radonc.2007.04.005. Epub 2007 May 9. Radiother Oncol. 2007. PMID: 17493699
-
Potential for reduced toxicity and dose escalation in the treatment of inoperable non-small-cell lung cancer: a comparison of intensity-modulated radiation therapy (IMRT), 3D conformal radiation, and elective nodal irradiation.Int J Radiat Oncol Biol Phys. 2003 Nov 1;57(3):875-90. doi: 10.1016/s0360-3016(03)00743-0. Int J Radiat Oncol Biol Phys. 2003. PMID: 14529795
-
Comparison of anisotropic aperture based intensity modulated radiotherapy with 3D-conformal radiotherapy for the treatment of large lung tumors.Radiother Oncol. 2012 Feb;102(2):268-73. doi: 10.1016/j.radonc.2011.10.006. Epub 2011 Nov 17. Radiother Oncol. 2012. PMID: 22100654
-
[Target volume definition, verification and correction in image-guided conformal intensity-modulated radiotherapy for non-small cell lung cancer].Zhonghua Zhong Liu Za Zhi. 2008 Apr;30(4):241-4. Zhonghua Zhong Liu Za Zhi. 2008. PMID: 18788623 Review. Chinese. No abstract available.
-
[Dosimetric comparison of different techniques for external beam accelerated partial breast irradiation].Magy Onkol. 2016 Nov 29;60(4):305-311. Epub 2016 Jun 15. Magy Onkol. 2016. PMID: 27898749 Review. Hungarian.
Cited by
-
Mid-position treatment strategy for locally advanced lung cancer: a dosimetric study.Br J Radiol. 2020 Jun;93(1110):20190692. doi: 10.1259/bjr.20190692. Epub 2020 Apr 15. Br J Radiol. 2020. PMID: 32293191 Free PMC article. Clinical Trial.
-
Treatment outcome and toxicity of intensity-modulated (chemo) radiotherapy in stage III non-small cell lung cancer patients.Radiat Oncol. 2012 Sep 7;7:150. doi: 10.1186/1748-717X-7-150. Radiat Oncol. 2012. PMID: 22958781 Free PMC article.
-
Variability of Low-Z Inhomogeneity Correction in IMRT/SBRT: A Multi-Institutional Collaborative Study.J Clin Med. 2023 Jan 23;12(3):906. doi: 10.3390/jcm12030906. J Clin Med. 2023. PMID: 36769553 Free PMC article.
-
Intensity-modulated radiation therapy (IMRT) for inoperable non-small cell lung cancer: the Memorial Sloan-Kettering Cancer Center (MSKCC) experience.Radiother Oncol. 2008 Apr;87(1):17-23. doi: 10.1016/j.radonc.2008.02.005. Epub 2008 Mar 17. Radiother Oncol. 2008. PMID: 18343515 Free PMC article.
-
Dosimetric impact of intermediate dose calculation for optimization convergence error.Oncotarget. 2016 Jun 21;7(25):37589-37598. doi: 10.18632/oncotarget.7743. Oncotarget. 2016. PMID: 26933998 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials