Measurement of pulsatile tidal volume, pressure amplitude, and gas flow during high-frequency percussive ventilation, with and without partial cuff deflation
- PMID: 17194317
Measurement of pulsatile tidal volume, pressure amplitude, and gas flow during high-frequency percussive ventilation, with and without partial cuff deflation
Abstract
Objective: With a high-frequency percussive ventilator and a mechanical lung model, to measure tidal volume (V(T)), pulsatile pressure amplitude (difference between peak and nadir pulsatile pressure [DeltaP];), and mean airway pressure (P (aw)) at various pulsatile frequencies, pulsatile inspiratory-expiratory ratios (I:E(p)), and pressures (measured at the interface between the pulse-generator and the endotracheal tube [P(vent)]).
Methods: With the endotracheal tube inside an artificial trachea, we manipulated the high-frequency percussive ventilation settings and adjuncts, including pulsatile frequency, I:E(p), and P(vent) by manipulating pulsatile flow. We also studied the effects of partially deflating the endotracheal tube cuff. We measured P (aw), pulsatile pressure amplitude at the carina (DeltaP(c)), and pulsatile V(T) at the carina. With the cuff partly deflated, we measured the fraction of inspired oxygen (F(IO(2))) in the gas efflux above and below the cuff.
Results: Increasing the pulsatile frequency from 300 cycles/min to 600 cycles/min and changing the I:E(p) from 1:3 to 1:1 significantly reduced V(T) (p < 0.001). P (aw) and DeltaP(c) were unaffected by the change in pulsatile frequency or I:E(p), except when we did not preserve the pulsatile flow. The measured V(T) range was from 19.1 mL (at 600 cycles/min) to 47.3 mL (at 300 cycles/min). Partial cuff deflation did not significantly reduce P (aw) or DeltaP(c), but it did significantly reduce V(T) and F(IO(2)).
Conclusion: During high-frequency percussive ventilation, the pulsatile frequency is inversely related to V(T). Partial cuff deflation reduces the delivered F(IO(2)).
Similar articles
-
Corrective measures for compromised oxygen delivery during endotracheal tube cuff deflation with high-frequency percussive ventilation.Respir Care. 2007 Mar;52(3):271-7. Respir Care. 2007. PMID: 17328825
-
Setting positive end-expiratory pressure during jet ventilation to replicate the mean airway pressure of oscillatory ventilation.Respir Care. 2007 Jan;52(1):50-5. Respir Care. 2007. PMID: 17194318
-
Mechanical loads modulate tidal volume and lung washout during high-frequency percussive ventilation.Respir Physiol Neurobiol. 2006 Jan 25;150(1):44-51. doi: 10.1016/j.resp.2005.02.015. Respir Physiol Neurobiol. 2006. PMID: 16448933
-
Tracheal gas insufflation and related techniques to introduce gas flow into the trachea.Respir Care. 2001 Feb;46(2):119-29. Respir Care. 2001. PMID: 11175241 Review.
-
A protocol for high-frequency oscillatory ventilation in adults: results from a roundtable discussion.Crit Care Med. 2007 Jul;35(7):1649-54. doi: 10.1097/01.CCM.0000269026.40739.2E. Crit Care Med. 2007. PMID: 17522576
Cited by
-
A narrative review of advanced ventilator modes in the pediatric intensive care unit.Transl Pediatr. 2021 Oct;10(10):2700-2719. doi: 10.21037/tp-20-332. Transl Pediatr. 2021. PMID: 34765495 Free PMC article. Review.
-
High-frequency percussive ventilation attenuates lung injury in a rabbit model of gastric juice aspiration.Intensive Care Med. 2008 Jan;34(1):91-100. doi: 10.1007/s00134-007-0848-z. Epub 2007 Sep 21. Intensive Care Med. 2008. PMID: 17885748