The performance of, and periprosthetic bone response to, a tapered, titanium (Ti6Al4V), hydroxyapatite-coated femoral hip prosthesis was evaluated at minimum of 10 years of follow-up. Data were prospectively collected on 147 consecutive primary hip arthroplasties performed in 133 patients by a single surgeon during a 2-year interval. Clinical and radiographic analyses of 96 hips in 86 patients were independently performed by 2 surgeons who were not involved in the care of these patients. There were no cases of aseptic loosening of the femoral component. Subsidence and stress shielding occurred in 5% and 2% of cases, respectively, and was not clinically significant. In all 15 hips that required revision of the acetabular component, the femoral component was found to be well fixed, without any occurrence of distal osteolysis. This femoral design provided reliable osseointegration that was durable at a mean of 11.5 years of follow-up.
Similar articles
-
[Long-term experience with the combined ARBOND hydroxyapatite coating in implant osteointegration].Acta Chir Orthop Traumatol Cech. 2009 Jun;76(3):172-8. Acta Chir Orthop Traumatol Cech. 2009. PMID: 19595277 Czech.
-
Total hip arthroplasty. Concerns with extensively porous coated femoral components.Clin Orthop Relat Res. 1998 Oct;(355):182-8. Clin Orthop Relat Res. 1998. PMID: 9917603
-
Seven-year results of a tapered, titanium, hydroxyapatite-coated cementless femoral stem in primary total hip arthroplasty.Clin Orthop Surg. 2010 Dec;2(4):214-20. doi: 10.4055/cios.2010.2.4.214. Epub 2010 Nov 5. Clin Orthop Surg. 2010. PMID: 21119937 Free PMC article.
-
Hydroxyapatite-coated external fixation pins.Expert Rev Med Devices. 2005 Jul;2(4):465-71. doi: 10.1586/17434440.2.4.465. Expert Rev Med Devices. 2005. PMID: 16293085 Review.
-
Osteoconductive coatings for total joint arthroplasty.Clin Orthop Relat Res. 2002 Feb;(395):53-65. doi: 10.1097/00003086-200202000-00007. Clin Orthop Relat Res. 2002. PMID: 11937866 Review.
Cited by
-
Is there any difference between tapered titanium stems with similar geometry and hydroxyapatite coating?Musculoskelet Surg. 2019 Dec;103(3):275-281. doi: 10.1007/s12306-018-00586-8. Epub 2019 Jan 8. Musculoskelet Surg. 2019. PMID: 30617813
-
Clinical application of artificial intelligence-assisted three-dimensional planning in direct anterior approach hip arthroplasty.Int Orthop. 2024 Mar;48(3):773-783. doi: 10.1007/s00264-023-06029-9. Epub 2023 Nov 15. Int Orthop. 2024. PMID: 37964157 Free PMC article. Clinical Trial.
-
Excellent midterm survival and functional outcomes of a fully hydroxyapatite-coated cementless stem: first results of a prospective multicenter study.Arthroplast Today. 2020 Mar 31;6(2):201-205. doi: 10.1016/j.artd.2020.01.009. eCollection 2020 Jun. Arthroplast Today. 2020. PMID: 32577462 Free PMC article.
-
10 years results of an uncemented metaphyseal fit modular stem in elderly patients.Indian J Orthop. 2011 Jul;45(4):351-8. doi: 10.4103/0019-5413.80328. Indian J Orthop. 2011. PMID: 21772630 Free PMC article.
-
RADIOGRAPHIC EVALUATION OF OSSEOINTEGRATION OF UNCEMENTED TARGOS® STEMS. A 5-YEAR FOLLOW-UP.Acta Ortop Bras. 2022 Dec 2;30(spe2):e250098. doi: 10.1590/1413-785220223002e250098. eCollection 2022. Acta Ortop Bras. 2022. PMID: 36506860 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical