Increased immunofluorescence sensitivity using 532 nm laser excitation
- PMID: 17200955
- DOI: 10.1002/cyto.a.20358
Increased immunofluorescence sensitivity using 532 nm laser excitation
Abstract
Objective: We evaluated the use of a high power, diode pulsed solid-state laser emitting 532 nm light for immunofluorescence applications. We compared the sensitivity and utility of this laser with the standard 488 nm excitation.
Methods: A flow cytometer was equipped with both a 488 nm and a 532 nm laser; fluorescence emissions from each laser were collected using the same filters and the same detector system. Cells or compensation beads (e.g. latex beads coated with anti-kappa antibodies) were stained with monoclonal antibodies conjugated to phycoerythrin (PE) as well as the PE tandem dyes TRPE, Cy5PE, Cy5.5PE, and Cy7PE. The sensitivity of detection of these reagents as well as those in heavily compensated channels was quantified by measuring the spreading error for a primary detector into a secondary detector.
Results: Measurement of the fluorescence emission of PE and PE-tandem dyes was considerably more sensitive when using 532 nm excitation (150 mW) as compared with 488 nm excitation (20 mW). In addition, as the absolute number of photoelectrons collected was greater, there was less measurement-error-induced spread into the compensated channels. As an example, when comparing the spreading error of PE labeled cells into the TRPE detector, the green laser was found to be 15-fold more sensitive as compared with the blue laser. In addition, the blue laser produced more autofluoresent signal from cells as compared with the green laser. Together, these advantages of the 532 nm excitation line provides for a significantly improved detection of immunofluorescence staining.
Similar articles
-
DPSS yellow-green 561-nm lasers for improved fluorochrome detection by flow cytometry.Cytometry A. 2005 Nov;68(1):36-44. doi: 10.1002/cyto.a.20182. Cytometry A. 2005. PMID: 16163703
-
Single laser three color immunofluorescence staining procedures based on energy transfer between phycoerythrin and cyanine 5.Cytometry. 1991;12(8):723-30. doi: 10.1002/cyto.990120806. Cytometry. 1991. PMID: 1794253
-
Phycoerythrin-allophycocyanin: a resonance energy transfer fluorochrome for immunofluorescence.Cytometry. 2001 May 1;44(1):24-9. Cytometry. 2001. PMID: 11309805
-
Immunofluorescence: quantitative considerations.Acta Histochem Suppl. 1988;35:77-83. Acta Histochem Suppl. 1988. PMID: 3138723 Review.
-
Detection of cytokine receptors by high-sensitivity immunofluorescence/flow cytometry.Immunobiology. 1992 Aug;185(2-4):350-65. doi: 10.1016/S0171-2985(11)80652-2. Immunobiology. 1992. PMID: 1452210 Review.
Cited by
-
Signal Improvement Strategies for Fluorescence Detection of Biomacromolecules.J Fluoresc. 2016 May;26(3):1131-9. doi: 10.1007/s10895-016-1806-3. Epub 2016 Apr 11. J Fluoresc. 2016. PMID: 27063869 Review.
-
Key steps and methods in the experimental design and data analysis of highly multi-parametric flow and mass cytometry.Comput Struct Biotechnol J. 2020 Mar 31;18:874-886. doi: 10.1016/j.csbj.2020.03.024. eCollection 2020. Comput Struct Biotechnol J. 2020. PMID: 32322369 Free PMC article. Review.
-
High-Dimensional Data Analysis Algorithms Yield Comparable Results for Mass Cytometry and Spectral Flow Cytometry Data.Cytometry A. 2020 Aug;97(8):824-831. doi: 10.1002/cyto.a.24016. Epub 2020 Apr 15. Cytometry A. 2020. PMID: 32293794 Free PMC article.
-
Near infrared lasers in flow cytometry.Methods. 2015 Jul 1;82:12-20. doi: 10.1016/j.ymeth.2015.03.010. Epub 2015 Mar 24. Methods. 2015. PMID: 25814439 Free PMC article.
-
Effect of a DPSS laser on the shear bond strength of ceramic brackets with different base designs.Lasers Med Sci. 2013 Nov;28(6):1461-6. doi: 10.1007/s10103-012-1227-8. Epub 2012 Nov 8. Lasers Med Sci. 2013. PMID: 23135786
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials