Clinical Trial
. 2006 Dec;29(12):1319-25.
doi: 10.1111/j.1540-8159.2006.00542.x.

ICD Implantation in infants and small children: the extracardiac technique

Affiliations
Clinical Trial

ICD Implantation in infants and small children: the extracardiac technique

Thomas Kriebel et al. Pacing Clin Electrophysiol. 2006 Dec.

Abstract

Background: There is no clear methodology for implantation of an internal cardioverter-defibrillator (ICD) in infants and small children. The aim of this study was to assess efficacy and safety of an extracardiac ICD implantation technique in pediatric patients.

Patients and methods: An extracardiac ICD system was implanted in eight patients (age: 0.3-8 years; body weight: 4-29 kg). Under fluoroscopic guidance a defibrillator lead was tunneled subcutaneously starting from the anterior axillar line along the course of the 6th rib until almost reaching the vertebral column. After a partial inferior sternotomy, bipolar steroid-eluting sensing and pacing leads were sutured to the atrial wall (n = 2) and to the anterior wall of the right ventricle (n = 8). The ICD device was implanted as "active can" in the upper abdomen. Sensing, pacing, and defibrillation thresholds (DFTs) as well as impedances were verified intraoperatively and 3 months later, respectively.

Results: In seven of eight patients, intraoperative DFT between subcutaneous lead and device was <15 J. In the eighth patient ICD implantation was technically not feasible due to a DFT >20 J. During follow-up (mean 14.5 months) appropriate and effective ICD discharges were noted in two patients. DFT remained stable after 3 months in four of six patients retested. A revision was required in one patient due to lead migration and in another patient due to a lead break.

Conclusions: In infants and small children, extracardiac ICD implantation was technically feasible. Experience and follow-up are still limited. The course of the DFT is unknown, facing further growth of the patients.

PubMed Disclaimer

Publication types