Correction of compliance and resistance altered by endotracheal tube leaks and non-linear pressure/volume-relationships
- PMID: 17206444
- DOI: 10.1007/s00421-006-0339-x
Correction of compliance and resistance altered by endotracheal tube leaks and non-linear pressure/volume-relationships
Abstract
Measurements of lung compliance (C) and resistance (R) are influenced by endotracheal tube leaks (ETTL) as well as non-linear pressure/volume relationships (P/V relationship). To keep C and R reliable, we developed an algorithm to mathematically correct inspiratory and expiratory volume (V) and flow. In this study, a ventilated lung model for non-linear P/V relationship with adjustment of an increasing ETTL was studied. In addition, the recordings (airway pressure, flow, and volume) of 21 infants (median weight: 1,220 g, range: 640-2,160 g, with a median leak size of 32%, range: 24-56%) were investigated. C and R were calculated continuously from the recordings of flow, volume, and airway pressure over time according to the changing volume. A method especially developed for the analysis of non-linear pressure-volume-relationship (APVNL) was employed. C and R affected by leaks were corrected applying the newly developed mathematical algorithm and compared with measurements without leakage. C could be corrected up to a leak of 80% and R up to 55% leak at half tidal V for the model with non-linear P/V-R. C and R without leak and after leak correction did not differ significantly in all infants where the APVNL method was applied (P > 0.05).
Similar articles
-
Correction of compliance and resistance altered by endotracheal tube leaks.Pediatr Crit Care Med. 2003 Jul;4(3):344-52. doi: 10.1097/01.PCC.0000059429.95034.73. Pediatr Crit Care Med. 2003. PMID: 12831418
-
Limitation of measurements of expiratory tidal volume and expiratory compliance under conditions of endotracheal tube leaks.Pediatr Crit Care Med. 2008 Jan;9(1):69-75. doi: 10.1097/01.PCC.0000298660.16328.BA. Pediatr Crit Care Med. 2008. PMID: 18477916
-
A new method to analyze lung compliance when pressure-volume relationship is nonlinear.Am J Respir Crit Care Med. 1998 Oct;158(4):1052-60. doi: 10.1164/ajrccm.158.4.9801011. Am J Respir Crit Care Med. 1998. PMID: 9769260
-
Role of lung function testing in the management of mechanically ventilated infants.Arch Dis Child Fetal Neonatal Ed. 2002 Jul;87(1):F7-F10. doi: 10.1136/fn.87.1.f7. Arch Dis Child Fetal Neonatal Ed. 2002. PMID: 12091280 Free PMC article. Review.
-
Building intelligent alarm systems by combining mathematical models and inductive machine learning techniques.Int J Biomed Comput. 1996 Apr;41(2):107-24. doi: 10.1016/0020-7101(95)01165-x. Int J Biomed Comput. 1996. PMID: 8803671 Review.
Cited by
-
Premature Infants Show Consistently Good Lung Compliance During Conventional Mechanical Ventilation.Pediatr Pulmonol. 2025 Jan;60(1):e27419. doi: 10.1002/ppul.27419. Epub 2024 Dec 2. Pediatr Pulmonol. 2025. PMID: 39620379 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials