Fast retrospectively gated quantitative four-dimensional (4D) cardiac micro computed tomography imaging of free-breathing mice
- PMID: 17220726
- DOI: 10.1097/01.rli.0000251572.56139.a3
Fast retrospectively gated quantitative four-dimensional (4D) cardiac micro computed tomography imaging of free-breathing mice
Abstract
Objective: We sought to demonstrate retrospectively gated dynamic 3D cardiac micro computed tomography (CT) of free-breathing mice.
Materials and methods: Five C57Bl6 mice were scanned using a cone-beam scanner with a slip-ring-mounted flat-panel detector. After the injection of an intravascular iodinated contrast agent, projection images were acquired over the course of 50 seconds, while the scanner rotated through 10 complete rotations. The mouse respiratory and electrocardiogram signals were recorder simultaneously with image acquisition. After acquisition, the projection images were retrospectively sorted into projections belonging to different cardiac time points, occurring only during expiration.
Results: Dynamic 3D cardiac images, with isotropic 150-microm voxel spacing, were reconstructed at 12-millisecond intervals throughout the cardiac cycle in all mice. The average ejection fraction and cardiac output were 58.2+/-4.6% and 11.4+/-1.3 mL/min, respectively. The measured entrance dose for the entire scan was 28 cGy. Repeat scans of the same animals showed that intrasubject variability was smaller than intersubject variability.
Conclusions: We have developed a high-resolution micro computed tomography method for evaluating the cardiac function and morphology of free-breathing mice in acquisition times shorter than 1 minute.
Similar articles
-
Prospective-gated cardiac micro-CT imaging of free-breathing mice using carbon nanotube field emission x-ray.Med Phys. 2010 Oct;37(10):5306-12. doi: 10.1118/1.3491806. Med Phys. 2010. PMID: 21089765 Free PMC article.
-
Optimization of a retrospective technique for respiratory-gated high speed micro-CT of free-breathing rodents.Phys Med Biol. 2007 Oct 7;52(19):5749-69. doi: 10.1088/0031-9155/52/19/002. Epub 2007 Sep 10. Phys Med Biol. 2007. PMID: 17881798
-
Micro-CT with respiratory and cardiac gating.Med Phys. 2004 Dec;31(12):3324-9. doi: 10.1118/1.1812604. Med Phys. 2004. PMID: 15651615 Free PMC article.
-
[New developments in computer tomography: EKG-correlated subsecond multi-slice spiral CT].Rontgenpraxis. 2001;53(4):134-40. Rontgenpraxis. 2001. PMID: 16231521 Review. German. No abstract available.
-
[Computed tomography of the lungs. A step into the fourth dimension].Radiologe. 2009 Aug;49(8):698-704. doi: 10.1007/s00117-009-1879-y. Radiologe. 2009. PMID: 19693618 Review. German.
Cited by
-
Left ventricle volume measurements in cardiac micro-CT: the impact of radiation dose and contrast agent.Comput Med Imaging Graph. 2008 Apr;32(3):239-50. doi: 10.1016/j.compmedimag.2007.12.004. Comput Med Imaging Graph. 2008. PMID: 18243656 Free PMC article.
-
4D micro-CT for cardiac and perfusion applications with view under sampling.Phys Med Biol. 2011 Jun 7;56(11):3351-69. doi: 10.1088/0031-9155/56/11/011. Epub 2011 May 10. Phys Med Biol. 2011. PMID: 21558587 Free PMC article.
-
Prospective-gated cardiac micro-CT imaging of free-breathing mice using carbon nanotube field emission x-ray.Med Phys. 2010 Oct;37(10):5306-12. doi: 10.1118/1.3491806. Med Phys. 2010. PMID: 21089765 Free PMC article.
-
A registration based approach for 4D cardiac micro-CT using combined prospective and retrospective gating.Med Phys. 2008 Apr;35(4):1170-9. doi: 10.1118/1.2868778. Med Phys. 2008. PMID: 18491508 Free PMC article.
-
Sparseness prior based iterative image reconstruction for retrospectively gated cardiac micro-CT.Med Phys. 2007 Nov;34(11):4476-83. doi: 10.1118/1.2795830. Med Phys. 2007. PMID: 18072512 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical