Objective measurement of postocclusion surge during phacoemulsification in human eye-bank eyes
- PMID: 17222793
- DOI: 10.1016/j.ajo.2006.11.017
Objective measurement of postocclusion surge during phacoemulsification in human eye-bank eyes
Abstract
Purpose: To objectively compare the postocclusion vacuum surge among different phacoemulsification machines and devices.
Design: Experimental study.
Methods: Infiniti, Legacy, Millennium, and Sovereign were tested in an eye-bank eye. All the machines were tested with 20-gauge non-ABS tips, 430 mm Hg vacuum pressure, 24 ml/minute aspiration rate, peristaltic pump, and 75 cm bottle height. In addition, Infiniti and Legacy were also tested with 20-gauge bypass tips (ABS), 125 cm bottle height, and 40 ml/minute flow rate. We also tested 19-gauge tips with Infiniti and Sovereign and the venturi pump for Millennium.
Results: Significant differences were found between all the machines tested with Millennium peristaltic generating the least and Millennium Venturi the most surge. ABS tips significantly decreased the surge for Legacy but not for Infiniti. Cruise Control (CC) had a significant effect on Sovereign but not on Millennium. Increasing the bottle height decreased surge while increasing the flow increased surge for both Infiniti and Legacy. The 19-gauge tips increased surge for both Infiniti and Sovereign.
Conclusions: Surge varied over a range of 40 microm to more than 2 mm. ABS and CC decrease surge, especially when the machine is not functioning near the limits of surge prevention. Certain parameters, such as a 19-gauge tip and high flow, dramatically increased surge, whereas elevating the bottle ameliorates it. Understanding the impact of all these features will help in minimizing the problem.
Similar articles
-
Efficacy of Cruise Control in controlling postocclusion surge with Legacy and Millennium venturi phacoemulsification machines.J Cataract Refract Surg. 2007 Jun;33(6):1071-5. doi: 10.1016/j.jcrs.2007.02.028. J Cataract Refract Surg. 2007. PMID: 17531704
-
Effect of bottle height and aspiration rate on postocclusion surge in Infiniti and Millennium peristaltic phacoemulsification machines.J Cataract Refract Surg. 2008 Aug;34(8):1400-2. doi: 10.1016/j.jcrs.2008.04.042. J Cataract Refract Surg. 2008. PMID: 18655995
-
A fluidics comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature phacoemulsification machines.Am J Ophthalmol. 2008 Jun;145(6):1014-1017. doi: 10.1016/j.ajo.2008.01.024. Epub 2008 Mar 17. Am J Ophthalmol. 2008. PMID: 18343350
-
Phacoemulsification in review: Optimization of cataract removal in an in vitro setting.Surv Ophthalmol. 2019 Nov-Dec;64(6):868-875. doi: 10.1016/j.survophthal.2019.06.007. Epub 2019 Jul 2. Surv Ophthalmol. 2019. PMID: 31276738 Review.
-
Innovations in phacoemulsification technology.Curr Opin Ophthalmol. 2007 Feb;18(1):23-6. doi: 10.1097/ICU.0b013e328011f9d0. Curr Opin Ophthalmol. 2007. PMID: 17159443 Review.
Cited by
-
Effect of Modified Peristaltic Phacoemulsification Device on Efficiency and Post-Occlusion Pressure Surge.Clin Ophthalmol. 2023 Aug 14;17:2341-2347. doi: 10.2147/OPTH.S416604. eCollection 2023. Clin Ophthalmol. 2023. PMID: 37600145 Free PMC article.
-
Comparison of occlusion break responses and vacuum rise times of phacoemulsification systems.BMC Ophthalmol. 2014 Jul 30;14:96. doi: 10.1186/1471-2415-14-96. BMC Ophthalmol. 2014. PMID: 25074069 Free PMC article.
-
Simulation of fluid dynamics and turbulence during phacoemulsification using the new propeller turbo tip.BMJ Open Ophthalmol. 2023 Sep;8(1):e001391. doi: 10.1136/bmjophth-2023-001391. BMJ Open Ophthalmol. 2023. PMID: 37775137 Free PMC article.
-
Optimization of the Oertli CataRhex 3® phacoemulsification machine.Clin Ophthalmol. 2019 Apr 16;13:633-639. doi: 10.2147/OPTH.S196281. eCollection 2019. Clin Ophthalmol. 2019. PMID: 31354230 Free PMC article.
-
Clinical Performance and Surgeon Acceptability of a New Dual Mode Phacoemulsification System.Clin Ophthalmol. 2022 Aug 6;16:2441-2451. doi: 10.2147/OPTH.S363061. eCollection 2022. Clin Ophthalmol. 2022. PMID: 35968054 Free PMC article. Clinical Trial.