The effect of contact lens induced corneal edema on Goldmann applanation tonometry measurements
- PMID: 17224766
- DOI: 10.1097/01.ijg.0000212277.95971.be
The effect of contact lens induced corneal edema on Goldmann applanation tonometry measurements
Abstract
Aim: To determine the effect of small increases in corneal hydration on the accuracy of Goldmann applanation tonometry estimates of intraocular pressure (IOP).
Materials and methods: Twenty-five young healthy subjects presented on 3 separate days approximately 1 week apart. On 2 visits, subjects were required to wear a hydrogel contact lens with either a center thickness of 0.3 and 0.7 mm (HEMA 38% water content, parallel surface curve) in 1 eye only under closed-eye conditions for 2 hours to induce corneal swelling. The third visit acted as a control. IOP, corneal thickness, and corneal curvature were measured in both eyes before and after contact lens wear on all visits.
Results: There was a statistically significant increase in corneal thickness of 40.2+/-14.4 microm (P<0.001) and 41.9+/-16.4 microm (P<0.001) after wearing the 0.3 and 0.7 mm thick contact lenses, respectively (2-tailed paired t test). There was an increase in IOP of 2.8+/-2.2 mm Hg (P<0.001) after wearing the 0.3 mm thick contact lens, and a statistically insignificant difference of 1.3+/-3.0 mm Hg (P=0.058) after wearing the 0.7 mm thick contact lens (2-tailed paired t test). There was a statistically significant Pearson correlation between the change in corneal thickness and the change in IOP after lens wear (r=0.500, P<0.001, 0.3 mm lens and r=0.399, P<0.001, 0.7 mm lens). The corneal hydration-induced measurement error was 0.46 mm Hg per 10 microm change in corneal thickness (0.3 mm lens) and 0.35 mm Hg per 10 microm change in corneal thickness (0.7 mm lens).
Conclusion: A small increase in corneal hydration and thickness may cause a clinically significant overestimation of IOP when measured using Goldmann applanation tonometry.
Similar articles
-
The effect of contact lens induced oedema on the accuracy of Goldmann tonometry in a mature population.Br J Ophthalmol. 2007 Dec;91(12):1636-8. doi: 10.1136/bjo.2007.118695. Epub 2007 May 15. Br J Ophthalmol. 2007. PMID: 17504854 Free PMC article.
-
The effect of corneal edema on dynamic contour and goldmann tonometry.Optom Vis Sci. 2008 Jun;85(6):451-6. doi: 10.1097/OPX.0b013e3181783a86. Optom Vis Sci. 2008. PMID: 18521023
-
The effect of contact lens-induced corneal edema on Goldmann applanation tonometry and dynamic contour tonometry.Graefes Arch Clin Exp Ophthalmol. 2009 Mar;247(3):371-5. doi: 10.1007/s00417-008-0954-9. Epub 2008 Oct 9. Graefes Arch Clin Exp Ophthalmol. 2009. PMID: 18843499
-
[The role of pachymetry in routine glaucoma diagnosis].Klin Monbl Augenheilkd. 2006 Feb;223(2):117-30. doi: 10.1055/s-2005-858724. Klin Monbl Augenheilkd. 2006. PMID: 16485223 Review. German.
-
Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.Surv Ophthalmol. 2000 Mar-Apr;44(5):367-408. doi: 10.1016/s0039-6257(00)00110-7. Surv Ophthalmol. 2000. PMID: 10734239 Review.
Cited by
-
Relationship Between Fuchs Endothelial Corneal Dystrophy Severity and Glaucoma and/or Ocular Hypertension.Arch Ophthalmol. 2012 Nov;130(11):1384-8. doi: 10.1001/archophthalmol.2012.1969. Arch Ophthalmol. 2012. PMID: 22777534 Free PMC article.
-
A retrospective analysis of intraocular pressure changes after cataract surgery with the use of prednisolone acetate 1% versus difluprednate 0.05%.Clin Ophthalmol. 2016 Nov 23;10:2329-2336. doi: 10.2147/OPTH.S121849. eCollection 2016. Clin Ophthalmol. 2016. PMID: 27920493 Free PMC article.
-
Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP.Graefes Arch Clin Exp Ophthalmol. 2010 Nov;248(11):1611-6. doi: 10.1007/s00417-010-1462-2. Epub 2010 Jul 22. Graefes Arch Clin Exp Ophthalmol. 2010. PMID: 20652303
-
Effects of aging on corneal biomechanical properties and their impact on 24-hour measurement of intraocular pressure.Am J Ophthalmol. 2008 Oct;146(4):567-572. doi: 10.1016/j.ajo.2008.05.026. Epub 2008 Jul 9. Am J Ophthalmol. 2008. PMID: 18614134 Free PMC article.
-
The effect of contact lens induced oedema on the accuracy of Goldmann tonometry in a mature population.Br J Ophthalmol. 2007 Dec;91(12):1636-8. doi: 10.1136/bjo.2007.118695. Epub 2007 May 15. Br J Ophthalmol. 2007. PMID: 17504854 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical