Accuracies of beta-lactam susceptibility test results for Pseudomonas aeruginosa with four automated systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2)
- PMID: 17229867
- PMCID: PMC1865858
- DOI: 10.1128/JCM.01716-06
Accuracies of beta-lactam susceptibility test results for Pseudomonas aeruginosa with four automated systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2)
Abstract
Contemporary clinical isolates and challenge strains of Pseudomonas aeruginosa were tested by four automated susceptibility testing systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2; two laboratories with each) against six broad-spectrum beta-lactams, and the results were compared to reference broth microdilution (BMD) and to consensus results from three validated methods (BMD, Etest [AB Biodisk, Solna, Sweden], and disk diffusion). Unacceptable levels of error (minor, major, and very major) were detected, some with systematic biases toward false susceptibility (piperacillin-tazobactam and imipenem) and others toward false resistance (aztreonam, cefepime, and ceftazidime). We encourage corrective action by the system manufacturers to address test biases, and we suggest that clinical laboratories using automated systems should consider accurate alternative methods for routine use.
References
-
- Biedenbach, D. J., and R. N. Jones. 1995. Interpretive errors using an automated system for the susceptibility testing of imipenem and aztreonam. Diagn. Microbiol. Infect. Dis. 21:57-60. - PubMed
-
- Biedenbach, D. J., S. A. Marshall, and R. N. Jones. 1999. Accuracy of cefepime antimicrobial susceptibility testing results for Pseudomonas aeruginosa tested on the MicroScan WalkAway system. Diagn. Microbiol. Infect. Dis. 33:305-307. - PubMed
-
- Burns, J. L., L. Saiman, S. Whittier, D. Larone, J. Krzewinski, Z. Liu, S. A. Marshall, and R. N. Jones. 2000. Comparison of agar diffusion methodologies for antimicrobial susceptibility testing of Pseudomonas aeruginosa isolates from cystic fibrosis patients. J. Clin. Microbiol. 38:1818-1822. - PMC - PubMed
-
- Burns, J. L., L. Saiman, S. Whittier, J. Krzewinski, Z. Liu, D. Larone, S. A. Marshall, and R. N. Jones. 2001. Comparison of two commercial systems (Vitek and MicroScan-WalkAway) for antimicrobial susceptibility testing of Pseudomonas aeruginosa isolates from cystic fibrosis patients. Diagn. Microbiol. Infect. Dis. 39:257-260. - PubMed
-
- Chandler, L. J., M. Poulter, B. Reisner, and G. Woods. 2002. Clinical evaluation of the Vitek automated system with cards GNS 122 and 127 and VTK-R07.01 software for antimicrobial susceptibility testing of Pseudomonas aeruginosa. Diagn. Microbiol. Infect. Dis. 42:71-73. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical