Daily electronic portal imaging of implanted gold seed fiducials in patients undergoing radiotherapy after radical prostatectomy
- PMID: 17236978
- DOI: 10.1016/j.ijrobp.2006.09.042
Daily electronic portal imaging of implanted gold seed fiducials in patients undergoing radiotherapy after radical prostatectomy
Abstract
Purpose: The aim of this study was to measure interfraction prostate bed motion, setup error, and total positioning error in 10 consecutive patients undergoing postprostatectomy radiotherapy.
Methods and materials: Daily image-guided target localization and alignment using electronic portal imaging of gold seed fiducials implanted into the prostate bed under transrectal ultrasound guidance was used in 10 patients undergoing adjuvant or salvage radiotherapy after prostatectomy. Prostate bed motion, setup error, and total positioning error were measured by analysis of gold seed fiducial location on the daily electronic portal images compared with the digitally reconstructed radiographs from the treatment-planning CT.
Results: Mean (+/- standard deviation) prostate bed motion was 0.3 +/- 0.9 mm, 0.4 +/- 2.4 mm, and -1.1 +/- 2.1 mm in the left-right (LR), superior-inferior (SI), and anterior-posterior (AP) axes, respectively. Mean set-up error was 0.1 +/- 4.5 mm, 1.1 +/- 3.9 mm, and -0.2 +/- 5.1 mm in the LR, SI, and AP axes, respectively. Mean total positioning error was 0.2 +/- 4.5 mm, 1.2 +/- 5.1 mm, and -0.3 +/- 4.5 mm in the LR, SI, and AP axes, respectively. Total positioning errors >5 mm occurred in 14.1%, 38.7%, and 28.2% of all fractions in the LR, SI, and AP axes, respectively. There was no significant migration of the gold marker seeds.
Conclusions: This study validates the use of daily image-guided target localization and alignment using electronic portal imaging of implanted gold seed fiducials as a valuable method to correct for interfraction target motion and to improve precision in the delivery of postprostatectomy radiotherapy.
Similar articles
-
Gold seed fiducials in analysis of linear and rotational displacement of the prostate bed.Radiother Oncol. 2014 Feb;110(2):256-60. doi: 10.1016/j.radonc.2013.10.037. Epub 2013 Dec 11. Radiother Oncol. 2014. PMID: 24332022
-
Prostate position relative to pelvic bony anatomy based on intraprostatic gold markers and electronic portal imaging.Int J Radiat Oncol Biol Phys. 2005 Nov 1;63(3):800-11. doi: 10.1016/j.ijrobp.2005.02.022. Int J Radiat Oncol Biol Phys. 2005. PMID: 16199313
-
Comparison of CT on rails with electronic portal imaging for positioning of prostate cancer patients with implanted fiducial markers.Int J Radiat Oncol Biol Phys. 2009 Jul 1;74(3):906-12. doi: 10.1016/j.ijrobp.2009.01.054. Int J Radiat Oncol Biol Phys. 2009. PMID: 19480970
-
Fiducial markers and spacers in prostate radiotherapy: current applications.BJU Int. 2014 Mar;113 Suppl 2:13-20. doi: 10.1111/bju.12624. BJU Int. 2014. PMID: 24894851 Review.
-
Image guided radiotherapy for prostate cancer.Bull Cancer. 2008 Mar;95(3):374-80. doi: 10.1684/bdc.2008.0599. Bull Cancer. 2008. PMID: 18390420 Review.
Cited by
-
Gold marker displacement due to needle insertion during HDR-brachytherapy for treatment of prostate cancer: a prospective cone beam computed tomography and kilovoltage on-board imaging (kV-OBI) study.Radiat Oncol. 2012 Feb 20;7:24. doi: 10.1186/1748-717X-7-24. Radiat Oncol. 2012. PMID: 22348595 Free PMC article.
-
Application of Hydrogel Spacer SpaceOAR Vue for Prostate Radiotherapy.Tomography. 2022 Oct 27;8(6):2648-2661. doi: 10.3390/tomography8060221. Tomography. 2022. PMID: 36412680 Free PMC article. Review.
-
Reducing errors in prostate tracking with an improved fiducial implantation protocol for CyberKnife based stereotactic body radiotherapy (SBRT).J Radiosurg SBRT. 2018;5(3):217-227. J Radiosurg SBRT. 2018. PMID: 29988326 Free PMC article.
-
Rationale and development of image-guided intensity-modulated radiotherapy post-prostatectomy: the present standard of care?Cancer Manag Res. 2015 Nov 11;7:331-44. doi: 10.2147/CMAR.S51955. eCollection 2015. Cancer Manag Res. 2015. PMID: 26635484 Free PMC article. Review.
-
Impact of the frequency of online verifications on the patient set-up accuracy and set-up margins.Radiat Oncol. 2011 Aug 24;6:101. doi: 10.1186/1748-717X-6-101. Radiat Oncol. 2011. PMID: 21864393 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical