. 2007 Mar;38(3):365-70.
doi: 10.1016/j.injury.2006.10.024. Epub 2007 Jan 22.

Locking compression plate with minimally invasive plate osteosynthesis in diaphyseal and distal tibial fracture: a retrospective study of 32 patients

Affiliations

Locking compression plate with minimally invasive plate osteosynthesis in diaphyseal and distal tibial fracture: a retrospective study of 32 patients

E Hasenboehler et al. Injury. 2007 Mar.

Abstract

With the introduction of Locking Compression Plates (LCP), Minimally Invasive Plate Osteosynthesis (MIPO) has become widely used. The plates act as internal fixators in a bridging manner, thus resulting in secondary bone healing. We retrospectively evaluated the healing pattern and the clinical evolution of diaphyseal and distal tibial shaft fractures over two and a half years in 32 patients (6 females, 26 males). Fractures were classified according to AO classification and included all 42A-C, 43A-B and 43C1-2 types. For open fractures, Gustillo Anderson classification was used. Plates consisted of the 4.5mm LCP and 3.5mm LCP-Pilon form plate. Clinical and radiological assessment was performed at 6 weeks, and at 3, 6, 9, and 12 months. Two patients were lost to follow-up. Fracture healing was defined as callus bridging of one cortex, seen on both lateral and posterior-anterior X-ray, and full, painless weight bearing. Ten patients at 3 months, 23 at 6 months, and 27 at 9 months met the criteria for a healed fracture. Plate bending was observed in one patient and called for re-operation at 5 months. Two patients required re-operation at 13 months secondary to pseudoarthrosis. Though MIPO seems more advantageous for soft tissue and bone biology, prolonged healing was observed in simple fracture patterns when a bridging plate technique was used.

PubMed Disclaimer