Effects on breast MRI of artifacts caused by metallic tissue marker clips
- PMID: 17242244
- DOI: 10.2214/AJR.05.1254
Effects on breast MRI of artifacts caused by metallic tissue marker clips
Abstract
Objective: The purpose of our study was to investigate MR artifacts related to tissue marker clips used in breast imaging procedures.
Materials and methods: Breast phantoms were created using gelatin doped with gadolinium. Four commercially available tissue marker clips were evaluated. Clinical MR evaluation of all phantoms with 1.5-T gradient-recalled echo sequences was performed. Images were evaluated for size and character of the visible artifacts and graphically appreciable fat saturation inhomogeneities. Quantitative measurement of the local inhomogeneity in 3D parts per million maps was obtained as a function of distance from each tissue marker.
Results: All tissue marker clips caused signal void artifacts on non-fat-suppressed images that measured 2-6 times the clip diameter. The degree of fat suppression inhomogeneity was minor but clinically appreciable. The local clip-induced field inhomogeneity varied from 0.25 to greater than 4.0 PPM for the four clips. At 0.25 PPM, the zonal diameter of frequency shift varied from 6 mm to 44 mm.
Conclusion: Artifacts caused by tissue marker clips could limit the sensitivity of MRI for detection and follow-up of breast cancer. The local effects on field inhomogeneity will affect local fat suppression and make spectroscopy data less reliable. These effects, though small, are measurable and vary among the different clips evaluated.
Similar articles
-
Comparison of the artifacts caused by metallic implants in breast MRI using dual-echo dixon versus conventional fat-suppression techniques.AJR Am J Roentgenol. 2014 Sep;203(3):W307-14. doi: 10.2214/AJR.13.10791. AJR Am J Roentgenol. 2014. PMID: 25148189
-
Titanium vs carbon coated ceramic breast tissue marker clips: 3T MR susceptibility artifact and local signal disturbance.Acad Radiol. 2011 Jun;18(6):770-3. doi: 10.1016/j.acra.2011.01.008. Epub 2011 Mar 21. Acad Radiol. 2011. PMID: 21419666
-
SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging.Invest Radiol. 2012 May;47(5):267-76. doi: 10.1097/RLI.0b013e318240a919. Invest Radiol. 2012. PMID: 22266987
-
Body MRI artifacts in clinical practice: a physicist's and radiologist's perspective.J Magn Reson Imaging. 2013 Aug;38(2):269-87. doi: 10.1002/jmri.24288. J Magn Reson Imaging. 2013. PMID: 23960007 Review.
-
Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT.Radiographics. 2007 May-Jun;27(3):791-803. doi: 10.1148/rg.273065087. Radiographics. 2007. PMID: 17495293 Review.
Cited by
-
Multispectral Imaging for Metallic Biopsy Marker Detection During MRI-Guided Breast Biopsy: A Feasibility Study for Clinical Translation.Front Oncol. 2021 Mar 22;11:605014. doi: 10.3389/fonc.2021.605014. eCollection 2021. Front Oncol. 2021. PMID: 33828972 Free PMC article.
-
Computational predictions of magnetic resonance acoustic radiation force imaging for breast cancer focused ultrasound therapy.Int J Hyperthermia. 2025 Dec;42(1):2452927. doi: 10.1080/02656736.2025.2452927. Epub 2025 Jan 22. Int J Hyperthermia. 2025. PMID: 39842813
-
Improvement in breast magnetic resonance imaging after a sentinel procedure for breast cancer with superparamagnetic tracers.Eur J Radiol Open. 2019 Jun 7;6:215-219. doi: 10.1016/j.ejro.2019.05.006. eCollection 2019. Eur J Radiol Open. 2019. PMID: 31198820 Free PMC article.
-
Agreement between MRI and pathologic breast tumor size after neoadjuvant chemotherapy, and comparison with alternative tests: individual patient data meta-analysis.BMC Cancer. 2015 Oct 8;15:662. doi: 10.1186/s12885-015-1664-4. BMC Cancer. 2015. PMID: 26449630 Free PMC article.
-
Quantification of breast biopsy clip marker artifact on routine breast MRI sequences: a phantom study.Eur Radiol Exp. 2024 Nov 15;8(1):128. doi: 10.1186/s41747-024-00525-2. Eur Radiol Exp. 2024. PMID: 39546114 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical