. 2007 Aug;17(8):1985-94.
doi: 10.1007/s00330-006-0564-z. Epub 2007 Feb 1.

Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom

Affiliations

Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom

Philipp G C Begemann et al. Eur Radiol. 2007 Aug.

Abstract

A calcium-scoring phantom with hydroxyapatite-filled cylindrical holes (0.5 to 4 mm) was used. High-resolution scans were performed for an accuracy baseline. The phantom was mounted to a moving heart phantom. Non-moving data with the implementation of an ECG-signal were acquired for different pitches (0.2/0.3), heart rates (60/80/95 bpm) and collimations (16 x 0.75/16 x 1.5 mm). Images were reconstructed with a cone-beam multi-cycle algorithm at a standard thickness/increment of 3 mm/1.5 mm and the thinnest possible thickness (0.8/0.4 and 2/1). Subsequently, ECG-gated moving calcium-scoring phantom data were acquired. The calcium volume and Agatston score were measured. The temporal resolution and reconstruction cycles were calculated. High-resolution scans determine the calcium volume with a high accuracy (mean overestimation, 0.8%). In the non-moving measurements, the volume underestimation ranged from about 6% (16 x 0.75 mm; 0.8/0.4 mm) to nearly 25% (16 x 1.5 mm; 3/1.5 mm). Moving scans showed increased measurement errors depending on the reconstructed RR interval, collimation, pitch, heart rate and gantry rotation time. Also, a correlation with the temporal resolution could be found. The reliability of calcium-scoring results can be improved with the use of a narrower collimation, a lower pitch and the reconstruction of thinner images, resulting in higher patient doses. The choice of the correct cardiac phase within the RR interval is essential to minimize measurement errors.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Biomed Tech (Berl). 2002;47 Suppl 1 Pt 1:243-5 - PubMed
    1. Eur Radiol. 2003 Mar;13(3):484-97 - PubMed
    1. AJR Am J Roentgenol. 2005 Jun;184(6):1917-23 - PubMed
    1. Acad Radiol. 2004 Mar;11(3):286-92 - PubMed
    1. Eur Radiol. 2002 Jun;12(6):1532-40 - PubMed

MeSH terms