[Ex-PRESS miniature glaucoma implant inserted under a scleral flap in open-angle glaucoma surgery: a retrospective study]
- PMID: 17287667
- DOI: 10.1016/s0181-5512(07)89545-3
[Ex-PRESS miniature glaucoma implant inserted under a scleral flap in open-angle glaucoma surgery: a retrospective study]
Abstract
Purpose: (1) To present the Ex-PRESS device implanted under a scleral flap in open-angle glaucoma patients and to report its safety and efficacy; (2) to compare the results obtained without a scleral flap in a glaucoma group.
Patients and methods: The case series studied included the eyes of Caucasian patients (82 patients, 99 eyes) with open-angle glaucoma, operated on between January 2003 and June 2004. The Ex-PRESS devices were inserted under the scleral flap in the anterior chamber; if necessary a combined surgery was performed (28 eyes). Each patient underwent ophthalmic examinations (IOP, visual field, gonioscopy, cup/disc ratio) before and after the operation. The surgical procedure lasted between 10 and 20 min with topical anesthesia. An antimetabolite was used under the scleral flap for all patients. The mean follow-up was 7.5+/-4.6 months; 40% had 12 months follow-up (40 eyes).
Results: The IOP decreased from 22.9+/-5.3 mmHg preoperatively to 14+/-2 mmHg at 6 months and 14.3+/-2.3 mmHg at 1 year. The success rate was 86.9% (IOP below 21 mmHg with or without drugs). Complete success was achieved in 62.6% (IOP below 21 mmHg without anti-glaucoma drops or medications). In 13 eyes, IOP was not controlled with eye drops, and nine eyes had to be reoperated. Six cases presented athalamia but recovered without surgical treatment. We did not observe any infection, corneal erosion, or Ex-PRESS extrusion. There were no statistical differences between results obtained with and without scleral flap regarding IOP or early complications (athalamia).
Conclusion: Both the safety and the efficacy of the device under a scleral flap were showed in glaucoma surgery in this retrospective study. Using the scleral flap reduces the risk of erosion but confirmation with a longer follow-up is needed.
Similar articles
-
Long-term outcomes and risk factors for failure with the EX-press glaucoma drainage device.Eye (Lond). 2014 Jan;28(1):1-8. doi: 10.1038/eye.2013.234. Epub 2013 Nov 15. Eye (Lond). 2014. PMID: 24232313 Free PMC article. Review.
-
Intermediate-term results of the Ex-PRESS miniature glaucoma implant under a scleral flap in previously operated eyes.Clin Exp Ophthalmol. 2011 Jul;39(5):421-8. doi: 10.1111/j.1442-9071.2010.02481.x. Epub 2011 Feb 23. Clin Exp Ophthalmol. 2011. PMID: 21176040
-
Implantation of a miniature glaucoma device under a scleral flap.J Glaucoma. 2005 Apr;14(2):98-102. doi: 10.1097/01.ijg.0000151688.34904.b7. J Glaucoma. 2005. PMID: 15741808
-
Phacoemulsification-deep sclerectomy modified by trabeculum microperforations and implantation of lens anterior capsule as autologous scleral implant.Curr Med Res Opin. 2010 Aug;26(8):2025-32. doi: 10.1185/03007995.2010.500857. Curr Med Res Opin. 2010. PMID: 20590394 Clinical Trial.
-
Long-term surgical outcomes of two different Ahmed Valve implantation techniques in refractory glaucoma: Scleral flap vs scleral tunnel.Eur J Ophthalmol. 2023 Jan;33(1):297-306. doi: 10.1177/11206721221097176. Epub 2022 Apr 27. Eur J Ophthalmol. 2023. PMID: 35473453 Review.
Cited by
-
Deep sclerectomy with the Ex-PRESS X-200 implant for the surgical treatment of glaucoma.Int Ophthalmol. 2010 Dec;30(6):661-8. doi: 10.1007/s10792-010-9382-z. Epub 2010 Jun 16. Int Ophthalmol. 2010. PMID: 20552258
-
Pars plana insertion of glaucoma shunt in eyes with refractory neovascular glaucoma: Case report.Medicine (Baltimore). 2018 Jun;97(23):e10977. doi: 10.1097/MD.0000000000010977. Medicine (Baltimore). 2018. PMID: 29879052 Free PMC article.
-
Surgical treatment of neovascular glaucoma with Ex-PRESS glaucoma shunt: Case report.Medicine (Baltimore). 2017 Sep;96(35):e7845. doi: 10.1097/MD.0000000000007845. Medicine (Baltimore). 2017. PMID: 28858093 Free PMC article.
-
Changes in anterior segment parameters following insertion of ExPRESS mini glaucoma implant vs. trabeculectomy.Arq Bras Oftalmol. 2020 Aug;83(4):283-288. doi: 10.5935/0004-2749.20200040. Epub 2020 Jul 29. Arq Bras Oftalmol. 2020. PMID: 32756788 Free PMC article.
-
Long-term outcomes and risk factors for failure with the EX-press glaucoma drainage device.Eye (Lond). 2014 Jan;28(1):1-8. doi: 10.1038/eye.2013.234. Epub 2013 Nov 15. Eye (Lond). 2014. PMID: 24232313 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources