Accuracy and limitations of exercise echocardiography in a routine clinical setting
- PMID: 1729348
- DOI: 10.1016/0735-1097(92)90054-q
Accuracy and limitations of exercise echocardiography in a routine clinical setting
Abstract
Despite the high reported accuracy of exercise echocardiography in the detection of coronary artery disease, factors that compromise its sensitivity and specificity are less clear. This study examined the results of 179 post-treadmill stress echocardiograms in 150 consecutive patients who also underwent cardiac catheterization and in 29 normal persons at low risk for coronary artery disease. Of 114 patients who had significant coronary stenoses at angiography, 96 had an abnormal exercise echocardiogram (overall sensitivity 84%). False negative results correlated with the performance of submaximal exercise, single-vessel disease and moderate (50% to 70% diameter) stenoses. After the exclusion of seven patients performing submaximal exercise, the sensitivity was 90%. In 54 patients without previous infarction performing maximal exercise, the sensitivity was 87%, higher in patients with multivessel coronary disease (96%) than in those with single-vessel disease (79%). After the exclusion of patients with nondiagnostic results, due either to the performance of submaximal stress or the presence of electrocardiographic (ECG) changes at rest, exercise echocardiography had a higher sensitivity than did exercise electrocardiography (87% vs. 63%, p = 0.01). In 36 patients without significant coronary disease, exercise echocardiography had an overall specificity of 86%. After the exclusion of patients with a nondiagnostic test, exercise echocardiography had a specificity of 82% compared with 74% specificity for exercise electrocardiography (p = NS). Similarly, of the 29 normal subjects, 93% had a normal exercise echocardiogram and 97% had a normal exercise ECG (p = NS). Similarly, of the 29 normal subjects, 93% had a normal exercise echocardiogram and 97% had a normal exercise ECG (p = NS). Age, gender, body weight and image quality did not significantly influence the accuracy of exercise echocardiography.(ABSTRACT TRUNCATED AT 250 WORDS)
Comment in
-
Exercise echocardiography: phase II, convincing the skeptics.J Am Coll Cardiol. 1992 Jan;19(1):82-3. doi: 10.1016/0735-1097(92)90055-r. J Am Coll Cardiol. 1992. PMID: 1729349 No abstract available.
Similar articles
-
Detection of coronary artery disease with upright bicycle exercise echocardiography.J Am Soc Echocardiogr. 1993 Mar-Apr;6(2):186-97. doi: 10.1016/s0894-7317(14)80489-6. J Am Soc Echocardiogr. 1993. PMID: 8481247
-
Diagnostic accuracy of peak exercise echocardiography in coronary artery disease: comparison with thallium-201 myocardial scintigraphy.Am Heart J. 1991 Dec;122(6):1609-16. doi: 10.1016/0002-8703(91)90278-p. Am Heart J. 1991. PMID: 1957756
-
Influence of left ventricular hypertrophy on detection of coronary artery disease using exercise echocardiography.J Am Coll Cardiol. 1995 Nov 1;26(5):1180-6. doi: 10.1016/0735-1097(96)81472-0. J Am Coll Cardiol. 1995. PMID: 7594030 Clinical Trial.
-
Digital supine bicycle stress echocardiography: a new technique for evaluating coronary artery disease.J Am Coll Cardiol. 1993 Mar 15;21(4):950-6. doi: 10.1016/0735-1097(93)90352-2. J Am Coll Cardiol. 1993. PMID: 8450164 Review.
-
Methodology, feasibility, safety and diagnostic accuracy of dobutamine stress echocardiography.J Am Coll Cardiol. 1997 Sep;30(3):595-606. doi: 10.1016/s0735-1097(97)00206-4. J Am Coll Cardiol. 1997. PMID: 9283514 Review.
Cited by
-
The relationship between cardiac hemodynamics and exercise tolerance in cystic fibrosis.Heart Lung. 2016 May-Jun;45(3):283-90. doi: 10.1016/j.hrtlng.2016.03.001. Epub 2016 Apr 1. Heart Lung. 2016. PMID: 27045901 Free PMC article.
-
Comparison of treadmill exercise stress cardiac MRI to stress echocardiography in healthy volunteers for adequacy of left ventricular endocardial wall visualization: A pilot study.J Magn Reson Imaging. 2014 May;39(5):1146-52. doi: 10.1002/jmri.24263. Epub 2013 Oct 10. J Magn Reson Imaging. 2014. PMID: 24123562 Free PMC article. Clinical Trial.
-
Dynamic stress echocardiography for evaluating anti-ischemic drug profiles in post-MI patients.Int J Card Imaging. 1997 Dec;13(6):485-91. doi: 10.1023/a:1005882829544. Int J Card Imaging. 1997. PMID: 9415850
-
Prognostic value of real time dobutamine stress myocardial contrast echocardiography in patients with chest pain syndrome.Int J Cardiovasc Imaging. 2011 Dec;27 Suppl 1(0 1):103-12. doi: 10.1007/s10554-011-9976-1. Epub 2011 Dec 6. Int J Cardiovasc Imaging. 2011. PMID: 22143170 Free PMC article.
-
Stress echocardiography: methods, indications and results.Neth Heart J. 2002 Mar;10(3):145-149. Neth Heart J. 2002. PMID: 25696080 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous