Prevention of dialysis hypotension episodes using fuzzy logic control system
- PMID: 17299006
- DOI: 10.1093/ndt/gfl799
Prevention of dialysis hypotension episodes using fuzzy logic control system
Erratum in
- Nephrol Dial Transplant. 2007 Jul;22(7):2104
Abstract
Background: Automatic systems for stabilizing blood pressure (BP) during dialysis are few and only control those variables indirectly related to BP. Due to complex BP regulation under dynamic dialysis conditions, BP itself appears to be the most consistent input parameter for a device addressed to preventing dialysis hypotension (DH).
Methods: An automatic system (ABPS, automatic blood pressure stabilization) for BP control by fluid removal feedback regulation is implemented on a dialysis machine (Dialog Advanced, Braun). A fuzzy logic (FL) control runs in the system, using instantaneous BP as the input variable governing the ultrafiltration rate (UFR) according to the BP trend. The system is user-friendly and just requires the input of two data: critical BP (individually defined as the possible level of DH risk) and the highest UFR applicable (percentage of the mean UFR). We evaluated this system's capacity to prevent DH in 55 RDT hypotension-prone patients. Sessions with (treatment A) and without (treatment B) ABPS were alternated one-by-one for 30 dialysis sessions per patient (674 with ABPS vs 698 without).
Results: Despite comparable treatment times and UF volumes, severe DH appeared in 8.3% of sessions in treatment A vs 13.8% in treatment B (-39%, P=0.01). Mild DH fell non-significantly (-12.3%). There was a similar percentage of sessions in which the planned body weight loss was not achieved and dialysis time was prolonged.
Conclusions: In conclusion, FL may be suited to interpreting and controlling the trend of a determined multi-variable parameter like BP. The medical knowledge of the patient and the consequent updating of input parameters depending on the patient's clinical conditions seem to be the main factors for obtaining optimal results.
Similar articles
-
[Effect of automated blood volume control on the incidence of intra-dialysis hypotension].Acta Med Croatica. 2003;57(1):17-22. Acta Med Croatica. 2003. PMID: 12876857 Croatian.
-
Oral intake during hemodialysis: is there an association with intradialytic hypotension?Hemodial Int. 2008 Jan;12(1):62-5. doi: 10.1111/j.1542-4758.2008.00242.x. Hemodial Int. 2008. PMID: 18271843
-
[How can we improve symptomatic hypotension in hemodialysis patients: cold dialysis vs isothermic dialysis].Nefrologia. 2007;27(6):737-41. Nefrologia. 2007. PMID: 18336104 Spanish.
-
Using dialysis machine technology to reduce intradialytic hypotension.Hemodial Int. 2011 Oct;15 Suppl 1:S37-42. doi: 10.1111/j.1542-4758.2011.00600.x. Hemodial Int. 2011. PMID: 22093599 Review.
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
Cited by
-
Fuzzy logic controller for hemodialysis machine based on human body model.J Med Signals Sens. 2011 Jan;1(1):36-48. J Med Signals Sens. 2011. PMID: 22606657 Free PMC article.
-
Development and validation of AI-based triage support algorithms for prevention of intradialytic hypotension.J Nephrol. 2023 Sep;36(7):2001-2011. doi: 10.1007/s40620-023-01741-6. Epub 2023 Sep 14. J Nephrol. 2023. PMID: 37707692
-
Closed Loop Ultrafiltration Feedback Control in Hemodialysis: A Narrative Review.Toxins (Basel). 2024 Aug 10;16(8):351. doi: 10.3390/toxins16080351. Toxins (Basel). 2024. PMID: 39195761 Free PMC article. Review.
-
Design and construct an optical device to determine relative blood volume in patients undergoing hemodialysis.Iran Red Crescent Med J. 2014 Apr;16(4):e15603. doi: 10.5812/ircmj.15603. Epub 2014 Apr 5. Iran Red Crescent Med J. 2014. PMID: 24910804 Free PMC article.
-
Feedback control in hemodialysis.Semin Dial. 2025 Jan-Feb;38(1):62-70. doi: 10.1111/sdi.13185. Epub 2023 Nov 23. Semin Dial. 2025. PMID: 37994191 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical